hemodynamic monitoring Flashcards
(82 cards)
3 aspects of hemodynamic monitoring
- arterial pressure
- central venous pressure
- central venous O2 sat (ScvO2)
principles of invasive monitoring nurses can:
- assess cardiac function
- circulating blood vol status
- physiological responses to medical and nursing interventions
invasive catheters 2
dependent on pressures
- ART line
- CVC
flush solution
NS
sometime heparin solution but can cause HIT
pressure bag
manual inflate to 300mmHg to ensure that blood from tubing doesnt go back up the tubing of pressure system
pressure tubing set
- is non-compliant to produce accurate reading
- allows continuous flow rate of 3ml/hr (under pressure)
- has fast flush device - allows bolus and flushes to clear blood during set up or to obtain blood sample
- typical- has 3 way stopclocks. one for blood sample, the other for zeroing
transducer
- senses BP in artery or vein
- BP transducer is translated into electrical signal to monitor
- also provides tracing (waveform) = BP and used for ECG monitoring
ECG
Electrocardiogram
why do you calibrate equipment? 2
to ensure accuracy of 2 baseline measurements
1. calibrate to atmospheric pressure “zeroing”
- determine the phlebostatic axis for transducer height placement “leveling the transducer”
leveling the transducer
- aligning the transducer with level of atrium
- line up air filled interface with the LEFT atrium to correct for changes in hydrostatic pressure in blood vessels above or below level of heart
- a carpenters level can be used to ensure phleb axis reference point
- if change in position, must do again
- if transducer is too high = false low BP reading
phlebostatic Axis
physical reference point on chest 4th intercoastal space to mid- axillary along with mid- anterior/posterior
- aprox level of atria
- can be pole mounted or arm mounted
- transducer must always be leveled to the phlebostatic axis
if the transducer is below the phleb axis…
we can think of it as the fluid in the system exerting extra weight on the transducer with reads as pressure inaccurate high readings
why do we calibrate equip?
to insure accuracy and 2 baseline measurements are needed
what is calibrating system to atmospheric pressure
zeroing
zeroing transducer
calibrate to atmospheric pressure
- 3 way stopcock nearest transducer is turned to open to air and close to PT and flush system
- monitor adjusts to zero (instead of atmospheric pressure which is 760mmHg)
- zero provides baseline
leveling step-by-step
- PT supine with HOB 0-60 degrees. doc HOB for reference
- locate the phlebostatic axis
- 4th intercost space
- axilla midline btwn anterior and posterior where X is - place carpenter level btwn phlebo axis and air filled interface (air reference stop cock) of transducer
- move transducer up/down IV pole until air-filled interface is centered
why do you position PT supine and reference HOB
to ensure accuracy of the readings
why locate the phlebo axis?
physical point of level of transducer reduces the effect of hydrostatic forces on transducer
ensures consistency of readings
why place carpenters level and move transducer up/down?
to ensure transducer air reference stopcock is leveled with the level of the right atrium
when will you see a flat waveform on the monitor
when you open the stopcock to air
why do you open to air?
the monitor can use atmospheric pressure as reference
steps to zeroing
- open stopcock to air- see flat waveform
- push and release zero button
- turn stopcock back to monitoring position and observe waveform
“open to PT, off to air”
where is the phlebo axis
level of the R atrium
what if the transducer is below the phlebo axis?
inaccurate high reading