According to ACR Mammography Accreditation standards, what is the minimum frequency for evaluating the compression thickness indicator by the technologist?
monthly
During ACR phantom scoring, what will result in a full point being awarded when viewing the fibers?
> = 8 mm long
During ACR phantom scoring, what will result in a full point being awarded when viewing the specks?
4-6 specks
During ACR phantom scoring, what will result in a full point being awarded when viewing the masses?
≤½ & >¾ border
Mammographers must perform phantom image evaluation at least
weekly
The MQSA does not require facilities to have
ACR accreditation
The average glandular dose delivered during a single craniocaudal view of an FDA-accepted phantom simulating a standard breast shall not exceed ____ mGy per exposure.
3
The mAs recorded during phantom image testing of mammographic equipment should not vary by more than ____ from the initial control value.
±15%
The maximum allowable dose per exposure for mammographic images is ____ rad.
0.3
The recommended maximum tolerance of kVp variance for mammographic x-ray generators is
5%
A visual inspection of a CT unit should be evaluated at least _____ for ACR accreditation programs.
monthly
For ACR CT accreditation, the Water CT Number and Standard Deviation (Noise) should be evaluated by the CT technologist
daily
For a slice thickness of 5 mm or less, the slice thickness should not vary by more than ____ mm.
0.5
Patient dose values for CT scans should remain within ____ of the manufacturer’s specifications.
±10%
Section thickness should not vary by more than ____ mm for sections greater than 5 mm.
±1.0
The CT number of water should not deviate by more than ____ CT numbers from 0.
3
The CTDIvol pass/fail criterion for an adult abdomen is ____ mGy.
30
The laser light field and radiation field should coincide to within ____ mm.
2.0
The minimum distance between two objects that allows them to be seen as separate and distinct would define what CT imaging parameter?
high contrast resolution
The uniformity of the CT number of water should not vary by more than ____ HU from the center of the phantom to the periphery
5
A change in the strength of the main magnetic field can be caused by
a. All are correct.
b. external ferromagnetic materials.
c. cryogen boiloff.
d. shim coil failure.
a. All are correct
An acquisition with a 200-mm field of view and 256 x and y encodes yields a pixel size of ____ mm2.
0.78
Luminance uniformity of soft-copy displays used in MRI systems should be within ____ of the maximum brightness measured in the center of the screen.
30%
Nonuniformity in the magnetic field can be caused by
a. All are correct.
b. the presence of ferromagnetic material within the bore of the magnet.
c. uniformity in the RF coil.
d. a homogenous main magnetic field.
b. the presence of ferromagnetic material within the bore of the magnet.