Image Production Flashcards
Which of the following reduces the scatter radiation reaching the detectors?
A. Post-patient collimation
B. Pre-patient collimation
C. Pre-detector collimation
D. Both a and c
D. Both a and c
Pre-detector collimation and post-patient collimation are synonymous. Both terms describe a device designed to remove scatter radiation before it reaches the detector.
Which of the following is the primary interaction between x-ray photons and tissue during computed tomography?
A. Bremsstrahlung
B. Characteristic
C. Compton effect
D. Coherent scatter
C. Compton effect
Owing to the high-energy beam used in CT, the Compton effect is the predominant interaction between x-ray and matter. CT scanners use between 120 and 140 kVp, yielding x-ray energies with averages between 70 and 80 keV.
The assignment of different generations to CT scanners is based on the configuration of the:
A. patient and gantry.
B. tube and detectors.
C. anode and cathode.
D. tube and collimators
B. tube and detectors
CT scanners are often placed into one of several different “generations.” The differences in generations are based on the relationship of the tube and detectors and the position of each during data acquisition.
When one is using a third-generation CT scanner, it is important that reference detectors positioned at the peripheral portions of the detector array be exposed to:
A. homogeneous radiation.
B. unattenuated radiation.
C. monochromatic radiation.
D. remnant radiation.
B. unattenuated radiation.
Reference detectors are used to measure incident radiation intensity. This information is used by the computer during the calculation of the linear attenuation coefficient. If the patient is placed incorrectly within the gantry, the reference detectors may be partially blocked, causing an out-of-field artifact.
In 1979, the scientists __________ and __________ shared the Nobel Prize for their research in computed tomography.
A. Watson and Crick
B. Olendorf and Hounsfield
C. Hounsfield and Cormack
D. Hounsfield and Ambrose
C. Hounsfield and Cormack
Dr. Godfrey Hounsfield and Dr. Allan Cormack shared the Nobel Prize for their work on the development of computed tomography in 1979.
The type of compensating filter used at the x-ray tube of a CT scanner is called a:
A. kernel.
B. water bath.
C. bow-tie filter.
D. wedge filter.
C. bow-tie filter.
The bow-tie filter used at the x-ray tube of a CT scanner absorbs a larger amount of radiation at the periphery of the patient, where the part thickness is reduced. The center of the patient is placed at the center of the filter, where the largest amount of radiation is allowed to pass through. The use of this type of filter attempts to compensate for the differences in thickness of the often oval-shaped patient.
The interaction between x-ray and matter that is responsible for the production of the scatter radiation absorbed by the patient and detectors is:
A. Compton effect.
B. bremsstrahlung.
C. photoelectric.
D. characteristic.
A. Compton effect.
In the Compton interaction, an x-ray photon ejects an outer-shell electron of an atom. The photon loses some of its energy in the collision and then continues on in a different, scattered direction. This interaction is the major source of the scatter radiation involved in the formation of the CT image.
The types of detectors used in computed tomography are:
- Gas ionization
- Stimulable phosphor
- Scintillation
A. 1 only.
B. 3 only.
C. 1 and 3 only.
D. 2 and 3 only.
C. 1 and 3 only.
The types of detectors used in computed tomography are scintillation and gas ionization detectors. Both operate by measuring the amount of transmitted radiation passed through the patient and transmitting this information to the computer for image reconstruction. Newer MDCT systems utilize solid-state scintillation detectors exclusively.
Which of the following is used in gas ionization CT detectors?
A. Neon
B. Xenon
C. Helium
D. Nitrogen
B. Xenon
Gases with high atomic numbers, such as xenon, are used in ionization-type CT detector systems. The high atomic number of the xenon gas (Z = 54) increases the incidence of interaction with x-ray photons, thus improving the efficiency of the detector.
Which of the following statements concerning the translate-rotate mode of CT data acquisition is/are correct?
- 360-degree circular detector arrays are used.
- Data are collected only during translation.
- Was used in first- and second-generation CT scanners.
A. 2 only
B. 3 only
C. 1 and 2 only
D. 2 and 3 only
D. 2 and 3 only
The first- and second-generation CT scanners used a translate-rotate method of data acquisition. The x-ray tube and detector(s) translated across the patient’s head, recording transmission measurements. The entire system then rotated 1 degree. This process of translation-rotation then repeated itself for a total of 180 degrees. First- and second-generation CT scanners used anywhere from 2 to 30 detectors.
Which of the following terms accurately describes the type of x-ray beam used in a third-generation CT scanner?
A. Pencil beam
B. Fan beam
C. Nutating beam
D. Electron beam
B. Fan beam
The third-generation CT scanner uses a fan beam projected onto a wide detector array. The third-generation CT x-ray beam is commonly mistaken for the “pencil beam” used with only two detectors in the first-generation CT scanner.
Which of the following technologic advances has led to the development of spiral/helical CT scanning?
- Slip-ring technology
- Electron beam technology
- High-efficiency x-ray tubes
A. 1 only
B. 1 and 2 only
C. 1 and 3 only
D. 1, 2, and 3
C. 1 and 3 only
Spiral/helical CT scanners are a recent development in computed tomography technology. The advents of slip-ring technology and improvements in x-ray tube design have made this innovation possible. Slip-rings have taken the place of the cumbersome cables previously used to transmit the CT signal and supply power to the tube and detectors. This development enables the tube and detectors to rotate continuously around the patient, acquiring data in the form of a helix. The longer exposure times of up to 60 seconds require extremely efficient x-ray tubes with enormous heat capacities.
For a given CT acquisition, the calculated mAs applied to each reconstructed slice may be referred to as the:
A. effective mAs.
B. peak mAs.
C. absorbed mAs.
D. constant mAs.
A. effective mAs.
Effective mAs describes the calculated mAs per acquired slice. The primary influencing factor for effective mAs is table speed, which is an important component controlled by the selected detector pitch. The detector pitch chosen by the CT operator directly affects the speed at which the patient travels through the gantry. The mAs divided by the pitch yields the effective mAs for each slice in a given CT acquisition.
The approximate inherent filtration of the CT x-ray amounts to an aluminum equivalent of:
A. 1.2 mm.
B. 3.0 mm.
C. 5.8 mm.
D. 10.0 mm.
B. 3.0 mm.
The tube housing, cooling oil, and so on constitute the inherent filtration, which amounts to approximately 3.0 mm of aluminum-equivalent filtration.
In multislice CT (MSCT), the beam pitch is equal to the table feed per rotation divided by the:
A. number of detectors.
B. pre-patient collimation.
C. total collimation.
D. number of channels.
C. total collimation.
The beam pitch, which is unique to multislice CT, is equal to the table feed per rotation divided by the total collimation. The total collimation is equal to the combined thickness of all of the sections that are simultaneously acquired with each gantry rotation.
An acquisition is made on a 4-slice MSCT system with a detector array of 16 x 1.25-mm detector elements along the z-axis. With a selected beam width of 20 mm, what beam pitch would result in the table moving 35.00 mm for each rotation of the gantry?
A. 1.0
B. 1.5
C. 1.75
D. 2.0
C. 1.75
The beam pitch for a given acquisition is equal to the table feed per rotation divided by the total collimation. The total collimation for this acquisition is equal to the total number of sections (detectors) multiplied by the detector dimension, or 16 multiplied by 1.25 mm. The beam pitch may be therefore calculated by dividing 35.00 by 20, or 1.75.
In an MSCT system, the detector array is composed of multiple rows of individual detector elements along the:
A. x-axis.
B. y-axis.
C. z-axis.
D. entire circumference of the gantry.
C. z-axis.
A multislice CT (MSCT) system utilizes a third-generation curvilinear detector array with multiple rows of individual detector elements segmented along the longitudinal axis, or z-axis.
The number, length, and organization of the individual detector elements in an MSCT system are referred to as the:
A. detector configuration.
B. array pitch.
C. CT generation.
D. detector pitch.
A
Detector configuration refers to the number, length, and organization of the individual detector elements in an MSCT system.
The type of MSCT detector array that contains midline narrow elements flanked by wider detectors is called a(n):
A. uniform matrix array.
B. adaptive array.
C. hybrid array.
D. stationary array.
C
A hybrid array consists of two detector sizes. The narrower detectors are positioned midline, flanked by the wider detectors. For example, the central rows of a 16-channel MDCT system are 0.625 mm with the remaining peripheral rows at 1.25 mm.
A 64-slice MSCT system employs an array of 64 detectors, each with a dimension of 0.625 mm. What beam collimation is required to expose the middle 32 detectors of the array to transmitted x-radiation?
A. 10 mm
B. 20 mm
C. 32 mm
D. 64 mm
B
The selected detector configuration for a given acquisition determines the necessary beam collimation. The selected 32 detectors, each with an individual dimension of 0.625 mm, would require a beam collimation of 20 mm to expose each to transmitted radiation (32 ✕ 0.625 = 20 mm).
The component of the DAS responsible for strengthening the signal emitted from a detector is termed the:
A. digital-to-analog converter (DAC).
B. preamplifier.
C. high-resolution comb.
D. bow-tie filter.
B
The weak electronic signal emitted from the detectors in response to the measurement of transmitted radiation must be strengthened before it is utilized for image reconstruction. The preamplifier is the component of the data acquisition system (DAS) responsible for amplifying the detector signal.
In 1917, Austrian mathematician __________ proved that it was possible to reconstruct a three-dimensional object from the infinite set of all of its projections.
A. Radon
B. Tsien
C. Bracewell
D. Cormack
A
J. Radon, an Austrian mathematician, was responsible for some of the earliest scientific research for the reconstruction principles used in computed tomography. In 1917 he proved that it was possible to build an image of an object through the use of an extremely large set of its projections.
Which of the following types of image reconstruction was used in the first prototype CT scanner?
A. Convolution method
B. Iterative technique
C. Fourier transform
D. Back-projection
B
The first-generation prototype CT scanner designed by Dr. Godfrey Hounsfield used an iterative form of image reconstruction.
CT images that have been reconstructed from a portion of the data acquisition process in the hopes of reducing patient motion artifacts are called:
A. dynamic images.
B. subtraction images.
C. segmented images.
D. filtered images.
C
CT images are usually constructed from transmission data acquired during a 360-degree rotation of the x-ray tube. A CT image can also be constructed from a portion of the data acquisition phase. For example, if a patient moves during the last third of a 2-second scan, an image can be constructed from the first 240 degrees of tube rotation, yielding an image free of motion. This process, referred to as segmentation, is a software capability of many modern CT scanners.























