Bipolar
Low amplitude, fractioned, high frequency EMG signal
100us
Below 0.2
Which muscle activity does the red electrode capture?
Styloglossus, transverse/vertical
**what ahi range is indicated for the genio therapy?
15-65
**what are the 3 main selection criteria for patients receiving genio
1) ahi 15-65
2) refused, not tolerated, or failed cpap
3) less than a 25% mixed and central apnea of the total apnea
What are the recommended electrodes for HGNS implant procedures?
Prass paired with insulation opening, bipolar, 12-18mm (18mm preferred)
When setting up a NIM system, what is the recommended rate and pulse width for the probe?
Rate: 7/s
Pulse: 100us
How many electrodes in total are there and where are they placed?
4 total, 2 pairs. One pair on each side of the Genioglossus muscle, and one pair on each side of the styloglussus/hyoglossus
When do you put the electrodes in the patient’s tongue ?
Before putting on the sterile drape, after intubation
How many mA should you use when identifying the HGN as a whole?
.5-.8
How many mA should you use when identifying inclusionary and exclusionary branches of HGN?
.15-.5