Lecture 3 Flashcards
(31 cards)
If you don’t have a confocal microscope how can you get a confocal quality image using a regular fluorescent microscope?
Deconvolve it!
What is deconvolution?
Deconvolution uses software that is calibrated to your microscope system to mathematically move out of focused light back in focus.
How do you calibrate deconvolution into your microscope?
To calibrate it to your microscope you image fluorescent beads as a control. Because you know the size of the beads, you can easily get that out of focus light back in focus.
How does Fluorescence recovery after photobleaching (FRAP) work?
Using fluorescently tagged proteins (Eg. EGFP-Actin) A small regions is hit with a laser for a prolonged period of time until there is no fluorescence remaining. After time that fluorescent area recovers. The time tells us the turnover (how fast it comes back) time of those proteins you were looking at.
What can Fluorescent Resonance Energy transfer (FRET) be used for?
We can take advtange of the emitted light from a fluorophore to excite a nearby fluorophore to see if the 2 molecules are closely associated using this method.
How does Fluorescent Resonance Energy transfer (FRET) work?
Use 2 fluorescent proteins.
Excite the first one.
It emits light that will excite the second one.
Then collect the light from the second one only.
Eg. CFP excited at 433/emits at 475nm.
Yfp excites at 475/emits at 530.
What are we always limited by?
Abbe limit of 0.2um.
How can we get past the optical resolution limit?
Use Super-Resolution Microscopy
What is structured Illumination (SIM)?
A grid is placed between the light path and the camera.
3 pictures are taken with the grid in 3 different positions.
Only the light that is in the same positions is kept, leaving a sharp image with little out of focused light.
Can get to 100 nm resolution (not bad.
Very slow because you need 3 pictures for every plane.
What is Stimulated Emission Depletion (STED)?
It works just like the point-scanning confocal, but has a Ring (AKA. a depletion beam) around the laser point.
This makes the excited area much smaller.
It takes all of the point, puts them together to build an image.
Resolution ~30nm.
What is the problem with Stimulated Emission Depletion (STED)?
Very slow!
What is Photo-Activated Localization Miscroscopy (PALM)?
Thousands of images are collected each with only a few molecules of a photo-activable fluorescnet protein (Eg. PA-GFP) excited.
These molecules are black to start off with, then fluorescence when activated.
The centre of the excited photons will give the location of the PA-GFP molecule itself. That is what is documented.
What is the problem with Photo-Activated Localization Miscroscopy (PALM)?
It is very very very very slow!!
There is 10nm resolution (essentially perfect resolution).
What is Stochastic Optical Resolution Microscopy (STORM)?
It is nearly identical to PALM, and it is really slow.
There is 20nm resolution (when it was invented), now down to ~5nm (phenomenal)!
Instead of using PA-GFP-type of molecules (so starting black and activating the colour)… like with PALM.
Lasers are used to excite a photoswitchable flurochrome to a dark state and turn on the fluorescence with 1 wavelength and off with another.
The centre positions of each molecular image is plotted base off of the photons emitted.
What does Stochastic mean?
Random
What is Super-resolution Microscopy (lattice light sheet)?
It uses a sheet of light to illuminate 1 entire plane of the sample being imaged in that same plane.
All planes are imaged very quickly (up to 1000 images/second).
Resolution ~ 150nm (so it is super-resolution, but there are better microscopes for resolving structures).
This enables live 3D super-resolution imaging in any direction you want.
How thick or thin are the tissue sections for imaging?
Most imaging happens with either transparent cells or thin tissue sections (5-10um) or transparent organisms.
What happens with thick samples?
Like a whole brain that you want to see, the distribution of the blood suplly or trace a single neutron.
OR
A kidney and you want to see a protein in a specific cell in the natural context of the brain (not ripped out of the organ).
What can you do?
TISSUE CLEARING
What is tissue clearing?
Why would you want to clear the tissue?
It can give you the 3D arrangment of whtver you’re looking at within huge volumes (100um-several cm)
Couldn’t you just use confocal or mulitphoton imaging?
Yes, but image quality gets progressively worse as the thickness increases. Also for thick samples, you have to use infrared wavelengths of light.
By clearing a sample you get rid of much of the matter that will generate out of focused light (the lipids are the problem) and can use light in the visible spectrum.
How do you “clear” a sample?
Example: Hydrogel embedding
Embed the samples in a hydrogel
Remove lipids in 8% SDS
Immerse in a clearing solution (FocusClear, 80% glyverol, there are many others)
End-up with a clear sample
Depending on the technique, it can take hours to months
Can cause the tissue to shrink or swell
Can easily be used for immunolocalization as there is no membrane hampering the permeability of the probes being used
What is electron microscopy?
Electron microscopy uses a high quality velocity beam of electrons to shoot at the same.
Resolution of 0.0005nm (the size of an atom) due to the short wavelength of the electrons
40,000 times better than a light microscope!
Usually the resolution is ~0.1nm for biological samples; 2000X better than a light microscope; 300X better than super-resolution
All under an ultrahigh vacuum because air can absorb electrons.
Electrons are contained within an electric field
This field condenses the beam of electron through the sample
What are the limitations to electron microscopy?
Can’t do live imaging
What are the 3 types of electron microscopy?
1) Transmission EM (and ImmunoEM)
2) Scanning EM
3) 3D EM tomography
What is Transmission EM?
Trasmission electron microscopt lets you see structures within cells.
Samples are fixed by cross-linking the proteins into position using fixatives (Paraformaldehyde/ Glutaraldehyde).
Embedded in a resin
Sectioned into ~70nm sections.
Stained with heavy metals