Lesson 2 (PRELIMS) Flashcards
(63 cards)
X-ray Mammography became clinically acceptable with the introduction of _____ as the target and filter (_____) and the dedicated, _____, _____ image receptor (_____).
molybdenum 1966
single- emulsion, screen-film 1972
Focal Spot Size: Tilting the x-ray tube to achieve an even smaller effective focal spot ensures imaging of the tissue next to the _____.
Chest wall
By _____, ______, emphasis on _____, _____ and _____ raised mammography to the level of excellence in breast imaging.
1990
grid technique
compression
high-frequency generators
automatic exposure control (AEC)
Components of Equipment
- Generator
- Support system
- X-ray tube
- Collimator or cone
- Breast compressor
- Anti-scatter holder
- Cassette holder
- Automatic Exposure Control (AEC)
- Phototimer
- Detector system
Generator:
- _____ of x-ray energy is required for good contrast.
- _____ tube rating.
- 20-40 kVp
- 3-10 kW
- All mammography imaging system incorporate _____ generators.
- Such a generator accepts a _____ input, which is rectified and _____ to produce a direct current (DC) voltage waveform.
- A maximum limit of _____ is standard for preventing excessive patient radiation dose.
High-Voltage generator
- high frequency
- single-phase; capacitor-smoothed
- 600 mAs
Target Composition: Mammographic x-ray tubes are manufactured with a _____, _____ or ______ target.
Tungsten (W) 74, Molybdenum (Mo) 42 or Rhodium (Rh) 45
Target Composition: The x-rays most useful for enhancing differential absorption in breast tissue and for maximizing radiographic contrast are those in the range of _____.
17 to 24 keV
Target Composition: Characteristic radiation produced by ____ and ____ are effective for breast imaging.
Mo and Rh
Target Composition: Mammographic x-ray tube have _____ anode
● Anode angle – _____ degrees
● Tube tilt – _____ degrees
●Source-to-image receptor distance – _____ cm
Rotating
- 0 to 16
- 6 to 24
- 60 to 65
Types of AEC devices (Phototimers): Refers to a key component within a _____, which is used to convert X-rays directly into electrical signals.
Solid State Diode
solid-state detector
The size of the _____ is an important characteristic of mammography x-ray tubes because of the higher demands for _____ resolution.
Focal spot
Spatial
Focal Spot Size: Imaging of microcalcifications requires _____ focal spots.
Small
Focal Spot Size: Mammography x-ray tubes usually have stated focal- spot sizes – large and small of _____ and _____, respectively
0.3 mm and 0.1 mm
Filtration: At low kVp used for mammography, it is important that the x-ray tube window not attenuate the x-ray beam significantly. Therefore dedicated mammography x-ray tubes have either a _____ window or a very thin _____ window.
beryllium (Z=4)
borosilicate glass
Filtration: Most mammography x-ray tubes have inherent filtration in the window of approximately _____.
0.1mm Al equivalent
Mammographic Technique Chart
Compressed Breast Thickness
0-2
3-4
5-6
7-8
7-8
Target Filter
Kilovolt Peak
Compressed Breast Thickness
0-2
3-4
5-6
7-8
7-8
Target Filter
0-2 > Mo-Mo
3-4 > Mo-Mo
5-6 > Mo-Rh
7-8 > Mo/Rh
7-8 > Rh/Rh
Kilovolt Peak
0-2 = 24
3-4 = 25, 26
5-6 = 28
7-8 = 32
7-8 = 30 (a)
Heel Effect: The conic shape of the breast requires that the radiation intensity near the chest wall must be _____ than that to the nipple side to ensure near uniform exposure of the image receptor.
- Positioning the _____ over the chest wall of the patient and the _____ over the anterior portion (nipple) achieves better uniformity of the transmitted x-rays through the breast.
Higher
Cathode
Anode
Two types of AEC devices are used
- Ionization Chamber
- Solid - State Diode.
Filtration: _____ is equivalent to 0.1 mm Al material (inherent filter).
1mm of Be
Target and Filter Composition: The filters used in mammography are based on the “_____” principle and attenuate the radiation above the _____ energy of the specific filter material, either _____ or _____
k edge
molybdenum or rhodium
Advantages of Vigorous Compression:
Effect: Immobilization of breast
Reduced motion blur
Advantages of Vigorous Compression:
Effect: Reduced scatter radiation
Improved contrast resolution
Compression:
- A compressed breast is of _____ thickness; therefore, the response of the image receptor is _____.
- Vigorous compression must be used in x-ray mammography.
- Compression improves _____ and _____ and _____ the patient radiation dose.
- It is achieved with a low attenuating _____ attached to a compression device.
- _____ (_____) of force is typically used.
- A flat, _____ paddle (not curved) provides a uniform density image.
- _____ to the breast support table
- more uniform
- spatial resolution and contrast resolution and reduces
- lexan paddle
- 10 to 20 Newtons (22 to 44 pounds)
- 90 degree
- Parallel