we need to top the reservoir up, it can obly be filled with purified water (AWD)
T or F
ture
If it’s still wet after sterilisation we can remove the moisture with a sterilised cloth.
T or F
false
Contamination was identified at inspection but it is alright to wipe it off as its getting sterilised.
T or F
false
As long as we clean them between times, we can transport our instrumentation in any available container.
T or F
false
Operator doesn’t need PPE if they are only using the Ultrasonic
T or F
false
As the Ultrasonic is a powerful process there should be no need to further process the instrumentation, processed through this equipment.
Tor F
false
The standard disinfection temperature banding within our washer-disinfector is
Select one:
a. 80 - 85
b. 85 - 90
c. 75 - 85
d. 65 - 75
e. 90- 95
e
90-95
identify stage A in AWD

flush
identify stage B in AWD

wash
identify stage C in AWD

rinse
identify stage D in AWD

disinfection
identify stage E in AWD

drying
Guidance documents are just there for reference, I don’t need to follow the directions within them
T or F
false
The first stage of a vacuum capable or Type - B steriliser is
Select one:
a. Equilibration
b. Steam penetration
c. Air removal
d. Negative pulsing
e. Heat up
f. Positive pulsing
c
air removal
A Bowie-Dick is carried out a daily basis, What is it testing for
Select one:
a. Temperature
b. Steam penetration
c. Air removal
d. Presence of air
e. Steam dryness
b.
steam penetration
What chemical would I use within an Ultrasonic Bath
Select one:
a. High alkaline
b. Low alkaline
c. Detergent
d. Enzymatic
e. pH neutral
d
enzymatic
Any kind of instrumentation can be processed in a Type - N steriliser
T or F
false
Staff were required to work in the LDU that have never worked there before. This isn’t a problem as they have a good idea what they are doing.
t or f
false
The WD cycle failed half-way through. The instruments should still be fine to use though
t or false
false
There was not enough chemical in the chemical reservoir and the WD cycle failed. The daily checks would have prevented this from happening
t or false
true
The WD chemical was replenished with a chemical that was found under the sink. This should be fine
t or false
false
The temperature at the prewash stage of the washer disinfector should be
<35 degrees C
The chemicals are added to AWD during
wash stage
dosing of chemicals should follow
manufacturer’s instructions