Test 4 - CH 12, 13 Flashcards

(148 cards)

1
Q

What is the primary result of Compton interactions?

A

scatter

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
2
Q

two major factors that affects the amount of scatter radiation produced and exiting the patient

A

volume of tissue irradiated
kVp

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
3
Q

the volume of tissue depends on what two things?

A

part thickness
x-ray beam field size

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
4
Q

what decreases the x-ray field size and the amount of tissue irradiated?

A

beam-restricting devices

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
5
Q

what is used to improve image quality by absorbing scatter that exits the patient

A

grids

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
6
Q

(T/F) grids reduce scatter in the patient

A

false
- only reduces the scatter that reaches the IR

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
7
Q

beam restriction limits what and reduces what?

A

limits patient exposure
- reduces amount of scatter radiation produced w/i patient

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
8
Q

what changes the shape and size of the primary beam?

A

beam-restricting devices

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
9
Q

increasing collimation does what to the field size?

A

decreases field size

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
10
Q

beam restriction does what to contrast?

A

increase

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
11
Q

for collimation, quantum noise increases so what has to be adjusted?

A

mAs

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
12
Q

what happens to quantum noise when field size increases?

A

decreases

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
13
Q

when going from a 14x17 IR to a 8x10, what should be done to mAs

A

double

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
14
Q

a flat piece of lead (diaphragm) that has a hole (aperture) in it and is placed directly below the x-ray tube

A

aperture diaphragm

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
15
Q

where is an aperture diaphragm placed?

A

under collimator

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
16
Q

what is caused from the aperture diaphragm because of its close proximity to the radiation source

A

large area of unsharpness on the edges

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
17
Q

an aperture diaphragm with an extended flange attached to it

A

cone or cylinder

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
18
Q

cones and cylinders are slid onto the tube where?

A

directly below the window

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
19
Q

cones and cylinders does what to sharpness around the image

A

limit

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
20
Q

which is more useful, cones or cylinders

A

cylinders

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
21
Q

(T/F) cones and cylinders are almost always made to produce a circular projected field

A

true

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
22
Q

what is considered the best beam-restricting device

A

collimators

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
23
Q

where is a collimator located? and consists of what?

A

below tube window
-lead shutters

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
24
Q

what are collimators equipped with?

A
  • white light source
  • mirror
  • measurement
  • crosshairs
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
25
what automatically limits the size and shape of the primary beams to the size and shape of the IR
automatic collimators - positive beam-limiting (PBL)
26
(T/F) with a PBL you are allowed to go larger than the IR
false
27
devices that have very thin lead strips with radiolucent interspaces, intended to absorb scatter emitted from the patient
radiographic grids
28
in grids, lead lines and interspace material is covered by what?
aluminum front and back panel
29
Grid ration formula
height / width of interspace
30
grid patterns
linear: lead lines run in 1 direction cross-hatched: lines run at right angles
31
lead line orientation: nonfocused and focused grid
nonfocused: lead lines run parallel focused: lead lines are angled to match divergence of beam
32
(T/F) nonfocused grids allow more transmitted photons to reach IR
false focused
33
the convergent point/line determines what?
the focal distance of a focused grid
34
distance between the grid and the convergent line or point
focal distance (grid radius)
35
what is the focal range
recommended range of SID that can be used with a focused grid
36
stationary are what type of grids and examples (2)
nonmoving -cassette : grid attached to IR permanently -cap: permanently mounted grid and allows the IR to slide behind it
37
(T/F) when grids are stationary it's possible to see grid lines on the radiographic image
true
38
example of moving/reciprocating grids
potter-bucky diaphragm: below table-top and above the tray that holds IR
39
grid motion is controlled by the what?
x-ray exposure switch
40
grids move slightly back and forth in a ______ direction over the IR during the ____ exposure
lateral entire
41
short-dimension grids run ___ to the long axis, long lead strips run ___ to the long axis
perpendicular parallel
42
a decrease in the number of transmitted photons that reach the IR because of some misalignment of the grid
grid cutoff
43
when does an upside-down focused grid cutoff occur?
when a grid is placed upside down on the IR -results in grid line going opposite the angle of divergence
44
off-level grid cutoff occurs when
when the x-ray beam is angled across the lead
45
off center grid cutoff occurs when
when the CR of the beam is not aligned with the center of a grid
46
off-focus grid cutoff occurs when
when using an SID outside of the recommended focal range
47
(T/F) off focus grid cutoff radiographically look different, whether it's being too close or too far away
false look the same
48
what is the Moiré effect, how is it caused?
zebra pattern artifact that can occur when a stationary grid is used during CR imaging, caused by the grid frequency being similar to the laser scanning frequency
49
grid selection involves consideration of what 3 things?
contrast improvement patient dose likelihood of grid cutoff
50
air gap technique is based on what?
increased OID
51
(T/F) placing a lead shield on the x-ray table absorbs scatter radiation
true
52
Does using a lead blocker decrease pt dose? why?
No, only absorbing scatter that already left the pt
53
what is AEC?
automatic exposure control -used to control the amount of radiation reaching the IR by terminating the length of the exposure
54
what factors does the tech set when using AEC?
kVp, mA, IR, grid
55
AEC is excellent at producing what when used properly?
consistent levels of exposure
56
How does AEC system know when to terminate the exposure?
uses radiation detectors that takes the radiation transmitted through the patient and converts it into an electrical signal
57
two types of AEC
phototimers: 1st gen ionization chambers: more common
58
How many radiation detectors are there and what is the techs role with them?
3 -tech chooses which ones to activate
59
Phototimer AEC devices uses what to convert light to electricity?
fluorescent screen and a device
60
a photomultiplier (PM) tube is a device that converts what?
visible light energy into electrical energy
61
a solid-state device that performs the same function
photodiode
62
with exit-type devices, what serves as the detectors?
light paddles coated with fluorescent material
63
with exit-type devices, how is the timer tripped?
when radiation interacts with the paddles it produces light, which is then converted to electricity
64
which is used more, exit or entrance type devices? examples of both?
entrance more used exit = phototimer entrance = ionization chamber
65
with entrance-type devices, how is the timer tripped?
when exposed to radiation, the air in the chambers becomes ionized, creating an electrical charge, it travels along the wire to the timer circuit
66
mAs readout / second readout is what?
the actual amount of mAs used for that image
67
the kVp level selected must be at least what?
at least the minimum kVp to penetrate the part
68
using higher kVp does what to exposure time and overall mAs needed?
decreases decreases
69
increasing mA on the control panel does what to exposure time?
decreases/shortens (opposite is true)
70
(T/F) minimum response time refers to the longest time that the system can produce?
false shortest time
71
minimum response time with modern AEC is what?
1 ms
72
(T/F) minimum response time is usally longer with AEC systems than with other types of radiographic timers
true
73
what refers to the maximum length of time the x-ray exposure will continue when using an AEC system?
Backup time
74
(T/F) backup time is a safety mechanism when an AEC fails or the equipment isn't used properly
true
75
what allows the radiographer to adjust the amount of preset radiation detection values?
exposure adjustment (density controls)
76
regularly using _______ settings to produce an acceptable image indicates that a problem exists
plus or minus exposure adjustment
77
what chambers/detectors do you choose?
ones that will be superimposed by the anatomic structures of greatest interest
78
failure to use the proper detectors could result in what?
underexposure or overexposure to the IR
79
(T/F) failure to select the correct bucky can result in the pt and IR being exposed to excessive radiation
true
80
improper centering may result in what?
underexposure/overexposure to the IR
81
(T/F) AEC size can be adjusted
false - fixed and can't be adjusted
82
what should you do if the detector combination is larger than the area of interest?
use a manual exposure technique
83
how does the AEC compensate for patient thickness?
if thicker, the exposure time will lengthen to reach the preset exposure
84
what may require greater technical consideration when using AEC?
excessive bowel gas destructive pathologic conditions postive contrast media
85
how is collimation a factor in AEC systems no collimation vs close collimation
- can cause the exposure to be terminated - could overexpose the pt
86
what does the AEC do when anatomically programmed techniques are used in conjunction with AEC?
- selects and displays manual expo factors - selects and displays AEC detectors to be used
87
what are essential procedures to maintain the proper functioning of the AEC
calibration and QC
88
how can AEC performance be monitored?
by imaging a homogenous pt equiv. phantom with additional thickness plates
89
what are preestablished guidelines used to select standardized manual / AEC exposure factors for each type of exam?
exposure technique charts
90
what 3 things do exposure technique charts do
produce consistent quality images reduces repeat radiograph studies reduce patient exposure
91
for technique charts to be effective, what should be established and what must operate properly?
quality standards radiographic system
92
devices that measure part thickness
caliper
93
where can you find the exact location for measuring part thickness?
technique chart
94
(T/F) the primary expo. factors of kVp and mAs and Ir type, grid ration are included regardless of the type of technique chart used?
true
95
two types of technique charts
variable - fixed mAs fixed kVp - variable mAs
96
chart that is based on the concept that kVp can be increased as the anatomic part size increases
variable kVp - fixed mAs
97
what is critical to the efficacy of the variable kVp chart?
accurate measurement of part thickness
98
(T/F) determination of the baseline kVp for each anatomic area has been standardized
false not been standardized
99
chart that uses the concept of selecting an optimal kVp that is required for the exam and adjusting the mAs for variations in part thickness
fixed kVp - variable mAs
100
what is kVp that is high enough to ensure penetration of the part but not too high to diminish radiographic contrast?
optimal kVp
101
(T/F) optimal kVp required for each anatomic area has not been standardized
true
102
once optimal kVp is established fixed kVp - variable mAs charts alter what?
the mAs for variations in part thickness
103
(T/F) using higher kVp ranges with DR systems can reduce the variability among exposure techniques for the same / similar regions
true
104
primary tools needed to develop a exposure technique chart
caliper phantoms calculator
105
this concept states that different parts of the same size can be imaged by use of the same expo factors, provided that the kVp needed to penetrate the part is used
comparative anatomy
106
(T/F) pediatric patients require high kVp and low mAs compared to adults
false low kVp
107
lower bone density may require what in exposure?
reduction
108
(T/F) geriatric patients may need an increase in exposure techniques
false
109
(T/F) geriatric patient may need an increase in kVp and mAs, grids, and imaged in quadrants
true
110
(T/F) diff projections and positions of the same part don't often require modification of exposure factors
false do require modification
111
fiberglass and plaster casts require what?
fiberglass = no change plaster = increase
112
inflatable, fiberglass, wood, aluminum, and solid plastic splints require what?
inflatable and fiberglass = no change everything else may be increased if it's in the path of the primary beam
113
diseases or conditions - that increases the absorption of the part - that decreases the absorption of the part
additive destructive
114
contrast media is used when anatomic tissues have what subject contrast?
low
115
contrast agent that - produces more brightness - produces less brightness
positive negative
116
When going from a 11x14 field size to a small 4in cone, what must be increased and why?
Increase exposure Need to compensate for the decrease in the number of X-ray photons that otherwise occur
117
Simplest type of beam restricting device
Aperture diaphragm
118
Can you adjust the field size with an aperture diaphragm?
No it’s a set size
119
Which beam restricting device has the least amount of area of unsharpness
Cones and cylinders
120
Grids are usually used when part thickness is what and kVp is higher than what?
Party thickness is greater than 10cm KVp greater than 60
121
Increasing grid ratio for the same grid frequency will increase what two things
Increase the amount of lead content Increase scatter absorption
122
If you increase grid frequency and use the same grid ratio, there is less what because why?
Less lead content - width of the interspace or thickness of the lead have been decreased
123
Decreasing the overall lead content will result in what?
Decreased scatter absorption
124
Which grid pattern absorbs more scatter and why
Crossed grids - contains more lead strips that are in two directions
125
Can you angle the X-ray tube when using a crossed grid
No Grid cutoff
126
Which type of grid (focused/non focused) allows more transmitted photons?
Focused
127
What grid matches the size of the cassette and is used by placing it on top of the IR
Wafer grid
128
How do the grid lines get blurred when using a stationary grid
Its slightly moved laterally (back and forth) during the entire X-ray exposure
129
GCF formula
mAs with grid / mAs without grid
130
What grid cutoff occurs with both focused and parallel grids
Off level - X-ray beam is angled
131
What can help eliminate the moire effect from happening
Higher grid frequency or moving grid
132
What describes the number of lead lines per unit length
Grid frequency
133
(T/F) difference in AEC systems lies in the type of device used to convert radiation into electricity
True
134
When the detectors are positioned behind the IR vs in front of the IR
Exit type device Entrance type device
135
Setting the backup time _________ of the expected exposure time is appropriate
150-200%
136
Is the detector able to distinguish transmitted radiation from scattered?
No
137
(T/F) IRs cannot be interchanged easily since the AEC is calibrated to terminate exposure levels at a preset level
True
138
What ensures the consistent and appropriate exposures to the IR are produced
Calibration
139
An X-ray generator should always have the same level of radiation exposure no matter what technical factors are set Max variable variation
Reproducibility of the exposure +- 5%
140
Variations between the state kVp and the X-ray beam quality must be within
+- 5%
141
QC checks used to determine kVp accuracy
Wisconsin test tool: holes in the IR Digital kVp meter
142
Exposure time directly affects the total quantity of radiation emitted from tube
Timer accuracy QC +- 5% for over 10ms +- 20% for less 10ms
143
The same mAs using different mA and time combinations should produce the same radiation output
Reciprocity law +- 10%
144
Sequential increases in mAs should produce the same sequential increases in exposure
Linearity +- 10%
145
Focal spot size variation Percent of blooming allowable for: 0.8 mm or less, 0.8mm to 1.5mm, 1.6mm or greater
50% 40% 30%
146
How well the collimator regulates the field size and area that is illuminated on the IR
Light field radiation fells congruence QC +- 2%
147
Center of the bucky tray must be centered to the X-ray beam
X-ray beam bucky alignment QC Within 1% of the SID
148
All medical X-ray units must be equipped with a SID indicator
SID indicator QC +- 2%