UV spec Flashcards
What makes up a UV spectrophometer?
1) Lamps – deuterium/tungsten
2) Monochromator
3) Optics
4) detector
How does the spectrophotometer work?
Polychromatic light enters a monochromator.
Monochromatic light leaves through exit slit with a selected wavelength and is made to pass through the sample.
The sample will absorb certain intensity of this light and the detector behind the sample records the light leaving the sample as a signal that is converted to a desirable format (Absorbance).
What happens in a dual beam instrument?
In dual beam instrument I o is split into two beams by mirrors and lenses. One of these pass through sample while the other passes through reference.
Reference compensates for absorption effects other than sample – e.g. solvents/buffers.
In the case of single beam the machine is zeroed to make this correction at the start.
How is the sample handled?
Sample is dissolved in a suitable solvent which should have minimum if any absorption in the λ range used.
e.g. for UV (>200nm): acetonitrile , ethanol, water.
What are cells (cuvettes) made out of?
plastic or glass for visible region but only quartz or silica is transparent in the UV region.
• Cuvettes handled from frosted side – filled to level above the beam. DO NOT NEED TO FILL COMPLETELY.
What is the ideal absorbance of the sample?
Ideally the absorbance of sample should be < 1.5 otherwise dilution is recommended
What can be done to compensate for the presence of extraneous materials in a drug sample?
Measure a difference spectrum (or absorbance) in drug sample before and after adjusting certain condition e.g. pH. Only works if the change selectively affects drug only.
What order is the plot of absorbance against wavelength?
ZERO ORDER
In derivative spectophotometry, what is the absorbance differentiated in respect to?
wavelength
dA/dλ = f ′(λ), first derivative d2A/dλ2 = f ″(λ), second derivative
What does differentiating do to the spectra?
Results in sharp peaks (rapid change in A vs λ) becoming more amplified. Gives a more characteristic profile.
Broad bands become less prominent compared to sharp bands. This effect increases with increasing order of the derivative.
NB: Derivatives simply help identify peaks it does not increase the data.
What is the process of luminescence?
(1) Excitation : electrons excited to higher energy state by light absorption
(2) Excited state life-time : ~ electrons stay in the excited state for a short time.
(3) Luminescence: Electrons return back to ground state and light is then emitted with less energy (some energy lost)
What is the Jablonski diagram?
Energy diagram that describes the process of photon emission is called Jablonski diagram
What is fluorescence?
return from excited singlet state to ground state; does not require change in spin orientation (more common of relaxation). The time spent from excitation back to So is called lifetime (τ) – typically few nanoseconds
What is internal conversion?
radiationless transition but vibrational levels need to match
What is intersystem crossing?
Molecules relax via a non-radiative transition to the T1
Requires spin orientation to change.
What is phosphorescence?
return from a triplet excited state to a ground state; electron requires change in spin orientation. Lifetimes are slow (msecs, second. Usually forbidden but it does happen. Phosphorescence has a longer lifetime than fluorescence (milliseconds rather than nanosecs). Phosphorescence generally occurs at longer wavelengths than fluorescence because the energy difference between S0 and T1 is lower
What is the process of fluorescence?
(1) Excitation – upon light absorption , a chromophore that was in lowest vibrational state (V0) of the ground state S0 is excited to some higher vibrational energy level (e.g. V1 or V2 etc) of S1 or S2
(2) Vibrational relaxations (non-radiative relaxation) takes place till the lowest (relaxed) vibrational level of S1 is reached. Molecule may undergo conformational change or interact with environment to achieve this state.
(3) The molecule then relaxes from this lowest vibrational energy level of the excited state S1 to one of the vibrational energy level of the ground S0 state. In doing so fluorescence is emitted.
takes about 10-15 s (2) 10-12 s, and (3) 10-9 s (or nanoseconds)
Describe the fluorescence spectrum
A plot of fluorescence emission against wavelength.
The spacings of the energy levels for the first excited electronic states are similar for absorption and fluorescence processes. This implies that the fluorescence spectrum of molecules can be near to mirror image of absorption spectrum but only for simple molecules.
What is Stokes shift?
The difference in wavelength between absorption and emission is called stokes shift.
as some of absorbed energy is lost due to processes that happen in the excited state lifetime. Examples of such processes :
Vibrational relaxation (heat), Collisional quenching, Energy transfer to other molecules (FRET), Intersystem crossing & Photobleaching
Hence the wavelength of light emitted by fluorescence is higher than the light that was absorbed.
Fluorescence excitation spectrum
for dilute solutions this is identical to absorption spectrum. It is a plot of iintensity of emission as a function of exciting wavelength. This spectrum helps to identify suitable wavelengths that can be used to excite the fluorophore.
Fluorescence emission spectrum
Fluorescence intensity against wavelength. The shape of this spectrum does not depend on which excitation wavelength was used whereas the intensity of emission does depend on the excitation wavelength.
What is more sensitive, Fluorescence or UV-Vis for detection of analytes?
FLUORESCENCE
Why is fluorescence more sensitive?
(1) The process is repeatable. Same molecule can be excited repeatedly thus producing many photons.
(2) Emission is at higher wavelength than absorption (stokes shift) so background signal can be quite low at detection wavelength. Thus better S/N ration
What is the instrumentation that measures fluorescence called and what does it contain?
Instruments that measures fluorescence is called spectrofluorometer
- ) Excitation source – Xe arc lamp, laser
- ) Cell holder- cells have to be four sided
- ) Monochromators
- ) Detectors
Laser –> Excitation monochromator –> emission monochromator –> detector