Week 1 Part 3 Flashcards

(52 cards)

1
Q

function of collimator

A

ensures gamma ray collection is from 1 orientation only

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
2
Q

components of gamma camera

A
  • collimator
  • photomultiplier tubes
  • arithmetic circuits
  • scintillation crystal of sodium iodide
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
3
Q

what are the outputs of arithmetric circuits to a CRO display

A

3-D coordinates for photon position

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
4
Q

arithmetic circuits combine ___ with that of its neighbour so that ___ for each point on the crystal can be obtained

A

output of each photomultiplier tube; electrical equivalent of light output

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
5
Q

size of scintillation crystal of sodium iodide

A

250 - 400mm

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
6
Q

function of photomultiplier

A

changes light into electrical energy

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
7
Q

What is NOT a type of collimator used in gamma cameras?

A

reflective

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
8
Q

shape of single detector gamma cameras

A

circular / rectangular

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
9
Q

how many detectors in mult-detectors

A

dual / triple

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
10
Q

types of dual detector systems

A
  • fixed and 180 degrees apart
  • fixed and 90 degrees apart
  • variable angles between 75 & 180 degrees apart
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
11
Q

Ri increases when crystal thickness ___ and gamma ray energy __

A

decreases; increases

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
12
Q

Rc is dependent on __

A

length of collimator holes & crystal thickness

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
13
Q

non uniformity is a measure of __

A

slightly different response of diff areas of detector to irradiation by uniform source

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
14
Q

intrinsic uniformity is dependent on __

A

PM tube balance, crystal, electronics

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
15
Q

extrinsic uniformity is dependent on __

A

PM tube balance, crystal, electronics and collimator integrity

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
16
Q

extrinsic uniformity is imaged with __

A

collimator using flood source

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
17
Q

intrinsic uniformity is imaged __

A

w/o collimator

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
18
Q

The purpose of the NaI(Tl) crystal is to

A

convert the gamma rays to light

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
19
Q

differential uniformity is the measurement of __

A

worst case rate of change of uniformity over a limited distance (~5 pixels)

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
20
Q

what are the typical integral & differential uniformity values after correction

A

1.5 - 3.5%; 1 - 2.5%

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
21
Q

What material is typically used to make collimators in gamma cameras?

22
Q

extrinsic uniformity test is performed with __

A

Co-57 flood source

23
Q

flood source is placed between ___

A

2 detector heads & acquired for 6,000,000 counts per detector

24
Q

energy resolution is ___

A

how well energy analyser of gamma camera is able to reject scattered photons and photons from other radioisotopes

25
analyser ensures that only gamma irradiation within ___ passes through
20% energy window
26
FWHM
full width at half maximum; measure of energy resolution & represents minimum difference necessary between energies of 2 gamma rays
27
if FWHM = 20keV, 2 gamma rays with a difference in energy of ___ cannot be distinguished by detector
< 20 KeV
28
the lower the FWHM, the __ the energy resolution
better
29
How would you select the correct window width for a particular radionuclide on a scintillation counter?
Plot a spectrum and set window width centred on photopeak and limits to exclude scatter and background.
30
count rate is the __
number of detected gamma photons per unit time
31
count rate characteristics refers to __
camera's response to varying count rate
32
gamma cameras are unable to work in high count rate environments due to ___
dead time
33
dead time refers to ___
time duration in micro seconds during which camera is unable to respond to another gamma irradiation
34
sensitivity refers to ___
detector's ability to use all gamma irradiation from patient efficiently
35
sensitivity can be improved by using a high sensitive collimator but __
SR with degrade
36
pulse height analysis and uniformity is tested __
daily
37
center of rotation is tested __
weekly
38
quadrant bar phantom followed by uniformity map creation is tested __
every 3 months
39
center of rotation test is used to ensure __
check accuracy of mechanical and electronic alignment of gamma camera's detector heads for accurate image reconstruction
40
any error in COR alignment leads to __
image artifacts or spatial distortion
41
What does a COR offset of more than the acceptable limit indicate?
The rotation axis does not match the detector center
42
what are the improved replacements for conventional PMTs
- position sensitive PMT - avalanche photodiode - silicon PMT
43
why are semiconductor detector precision of gamma signal detection better ___
no need for light electrical signal conversion = improved energy resolution
44
CdTe
cadmium telluride
45
CZT
cadmium zinc telluride
46
CZT and CdTe are better than HPGe in terms of __
energy resolution, better scatter rejection, compact and rugged package
47
In solid-state gamma cameras, what is the role of CZT detectors?
Convert gamma photons directly to electrical signals
48
What is commonly used in solid-state gamma cameras?
CZT
49
What is the main advantage of a solid-state gamma camera over an Anger-type gamma camera?
Improved spatial resolution and sensitivity
50
A consequence of attenuation in NM imaging is the improved appearance of ___
deep lesions.
51
Scintillation is a process that results in an emission of beta particles by a substance (scintillator) when it is bombarded by gamma rays. True or False
False
52
A disadvantage of sodium Iodide (Nal) crystal is that it is ___
hygroscopic & fragile.