Week 10: Fluoro Unit Configurations and Digital Acquisition Flashcards
(32 cards)
Image Intensifier (II)
- electronic device which receives the image forming x-ray beam
- purpose is to intensify and convert into visible light image
- newer inits replaced by flat panel detector (FPD)
Video Monitoring System (Mobile Viewing Station)
- flat panel monitors
- high resolution
- used for viewing images during the exam
- receives electronic signal and converts into a visible image
Fluoro Tube/Generator
- very similar to gen rad tubes
- converts electrical energy into x-rays
- designed to operate for longer exposures with lower mA
- generators are high frequency and higher heat capacity than gen rad
Tube Under Table Configuration
- allows operator to be close to patient during exam
- can be controlled using foot switch
- minimum source to skin distance is 38cm
- most common configuration
Tube Over Table Configuration
- greater distance between the tube and the patient and II which improves image quality and reducing radiation skin dose to patient
- radiation hazards to operator because of scatter but also from the primary beam
- tech can’t be as close to patient to be of assistance, must step back during exposure
C-Arm Configuration
- mobile
- can be positioned under or over the table
- commonly used in OR
- operators must always be vigilant of source to skin distance, minimum of 30cm
Housing
- metal construction, lead lines to absorb off-focus radiation
- provides structural support and protection from rough handling
- protects the vacuum
Angio Fluoroscopy Units
- large diameter, thick anodes because the x-ray tube must tolerate high heat loads
- high power rating for rapid sequence serial radiography
- c-arms are floor/ceiling mounted, table floats
- small (0.3mm) focal spots must be available for imaging small vessels
Glass Envelope
- vacuum tube that allows free flow of electrons from the photocathode to the anode
- can experience as much a 1 coin fro the outside
Input Phosphor Screen
- receives x-ray photons and emits light proportional to the absorption of the x-ray photons
- concave surface to improve spatial resolution by maintaining the distance between each point on the input screen and its corresponding location on the output screen
- coated with sodium activated cesium iodine (CsI) phosphor which emits light
- diameter 6 inches - 23 inches
Photocathode
- directly connected to the input screen, made of cesium antimony
- receives light and releases electrons by photoemission (directly proportional)
- an extremely thin protective coating is applied between the input screen and the photocathode
Electrostatic Lens
- series of charged electrodes inside the glass envelope
- negatively charged to accelerant, converge and focus electrons to the centre off the ring shaped anode
- elections are focused to a point called the focal point (reverses image)
Anode
- circular plate with. hole in the middle for electrons to pass though
- accelerates the electrons from the photocathode to the output phosphor/screen
- positively charged and located inside the envelope immediately in front of the output screen
- voltage difference between cathode and anode of 25-35 kV
Output Phosphor Screen
- made of thin layer of zinc cadmium sulphide, 1 inch in diameter
- converts electrons into light
- image at this point is much brighter due to smaller size and additional energy through acceleration process
- an opaque filter of aluminum is placed along the inside of the output screen to prevent light from returning to input screen
Charged Coupled Device (CCD)
Light sensitive semiconducting device that generates an electrical charge when stimulated by light and stores this charge in a capacitor, Reads out line by line, quantified and plotted on the matrix.
CCD Advantages
- more durable
- more sensitive to light (higher DQE)
- requires less radiation
- higher spacial resolution
- fast discharge time = good for high speed imaging like cardiovascular studies
Flat Panel Detector
- replacing the II
- utilizes indirect capture detector or direct capture detector (like gen rad)
Advantages of FPD
- streamlined assembly (reduced size and weight)
- large FOV
- multiple operational modes
- not effected by distortion
- less radiation is mag mode
- broader TFT dynamic range and low contrast detectability
Last Image Hold
- dose reduction technique during acquisition
- eliminates the need for an additional exposure
- last image from the exposure is help on the screen even though there is no active fluoro happening
Frame Averaging
- dose reduction technique during acquisition
- combining multiple frames to reduce noise, requires less mA to minimize QM
- objects must be moving slow or lag will be noticeable
Pulsed Fluoroscopy
- reduces dose to patients and healthcare providers while maintaining high quality images
- uses higher mA so there will be less QM resulting in better image sharpness
- the dose is pulsed so even though the mA is higher, the radiation is not being emitted constantly which reduces overall exposure
Interrogation Time
The time required for the tube to switch on and to get to selected mA and kV
Extinction Time
The time required for the x-ray tube to be switched off
Duty Cycle
The time the tube is energized. Ex: tube is energized for 100ms every second = 10% duty cycle. Decreased number of pulses per second = lower dose to patient and staff.