Week 5: CT Dose and QC Flashcards

(33 cards)

1
Q

Dose Uniformity

A

Decreases as SFOV and patient thickness increase.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
2
Q

Multiple Scan Average Dose (MSAD)

A
  • average dose patient gets for exam (sequential or helical)
  • includes central slice dose reading and “tails”
  • dose for entire examination
  • tissue dose, not just skin dose
  • Grays or mGys
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
2
Q

CT Dose Index (CTDI)

A
  • estimate of MSAD manufacturers use
  • dose measurements are taken on a phantom then programmed into the machine so that a tech knows about how much dose will be received when certain parameters are set.
  • used for protocol planning
  • done with contiguous slices, no gaps, no overlaps
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
3
Q

CTDI 100

A
  • CTDI created using 100mm pencil ionization chamber (to cover and therefore evaluate multiple slices at one) is used as the baseline
  • must take readings in different locations in the phantom
  • must take reading at each selectable condition that caries the rate or duration of irradiation
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
3
Q

CTDI w

A
  • weighted CT dose index
  • weighted or average CTDI 100 across entire single slice/section
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
4
Q

CTDI Vol

A
  • CTDI vol = CTDI w/pitch
  • take into account exposure in the z-axis
  • measure of exposure per slice independent of scan length
  • the CTDI vol is “fixed” and independent of patient size and scan length thus the CTDI volume does not quantify how much radiation any specific patient receives but simply indicates the intensity of the radiation being directed at the patient.
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
5
Q

Effective Dose

A
  • with DLP we know how much dose has been given but not how it affects individual tissues, effective dose considered tissue sensitivity
    WT = 0.12: stomach, colon, lung, red bone marrow, breast, remainder tissues
    WT = 0.08: gonads
    WT = 0.04: urinary bladder, esophagus, liver, thyroid
    WT = 0.01: bone surface, skin, brain, salivary glands.
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
5
Q

DLP

A
  • the total amount of radiation incident on the patient
  • the product of the CTDI vol and scan length (in cm)
  • the DLP is the second dose metric that is easily accessible to the radiologist and account for both radiation intensity (CTDI vol) and scan length of the CT exam
  • mGy x cm
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
6
Q

mA: Balance QM and Dose

A

Increasing mA will increase dose so use the lowest mA and time that will give acceptable SNR

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
7
Q

kV and Dose

A

kV changes have a massive effect on dose, 15% reduction is folate results in 35% dose reduction. Smaller patients and children must rescanned at lower kV.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
8
Q

Rotation Time affects Total Time

A

When it takes longer for the tube to make a rotation, dose is increased. When the tube rotates slower, spacial resolution is increased. Rotation time is inversely related to rotation speed.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
9
Q

Pitch and Dose

A

As pitch increases, dose decreases but so does spacial resolution. We must minimize overlap but be wary of pitches greater than one because data will be skipped and missed.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
9
Q

Thin Slices and Dose

A

Thin slices are good for detail but increase dose. We have to think about how thick we can acquire our slices and still get a good detail in reconstructed MPRs. Slice thickness increases = dose, noise and spacial resolution decreases.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
10
Q

Air Calibrations

A

Usually initiated with hr warm up cycle. HUs are real attenuation measurements so it must be assured every day that the detectors are returning accurate readings. Takes exposure at various settings (kV, SFOV, slice thickness, helical mode etc) and evaluates the CT numbers to see if they all return an HU appropriate for air. Corrects for detector drift/calibration errors.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
10
Q

CT Tube Warmup

A

Performed daily by the first tech on duty. Unit is turned on, unit generally initiates a warm up upon start up and takes many exposures to bring focal spot to appropriate heat.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
11
Q

CT HU Linearity

A

SC35 recommends this be done monthly, Toshiba recommends daily using one kV. Determines the CT numbers for certain know material densities. Uses phantom containing a variety of know densities.

12
Q

How to perform HU Linearity Test

A
  1. place the phantom on the couch with the holder
  2. scan the phantom
  3. place an ROI over each section on the resultant image
  4. record average HU
  5. scanner HU tolerances are used
13
Q

CT Number Accuracy

A

Determines accuracy of CT number for water. SC35 requires this dest be done weekly.

14
Q

How to perform CT Number Accuracy

A
  1. scan homogenous/uniform water phantom
  2. place an ROI near centre of image
  3. record the output HU
  4. must be within +/- 4 HU of zero
14
Q

CT Noise

A

Determines the standard deviation of CT numbers across an image. SC35 requires this be done weekly.

15
Q

How to perform CT Noise Test

A
  1. scan the uniform water phantom
  2. read ROI in one area and evaluate standard deviation within area
  3. must be within 10% or 0.2 HU
16
Q

CT Uniformity

A

Determines whether CT numbers are the same regardless of area of the image in a uniform phantom. SC35 requires this test be done weekly.

17
Q

How to perform CT Uniformity Test

A
  1. scan homogenous water phantom
  2. drop a ROI central and four peripheral areas
  3. must be within +/- 2 HU
  4. baseline values must be within +/- 5 HU from manufacturer
18
Q

Calibration of CT Number

A

Determines the mean CT number and standard deviation for water and air on a reconstructed image. SC35 requires this be done monthly. Uses water filled phantom, image and measure CT numbers for air/water with ROIs. CT number for water must be 0 +/- 4 HU. CT number for air must be -1000 +/- 10 HU.

19
CT Tomographic Section Thickness
Determines the section/slice thickness accuracy. SC35 requires this test be done monthly. Uses test device with ramp, spiral, step wedge for step and shoot or dick, beads or lines for helical scanning. Check outer and inner section areas.
20
CT Spatial Resolution /High Contrast Resolution
SC35 requires this test be done quarterly. Determines scanner speciation resolution. SC35 stated best method uses MTF with point spread function, thin contrast wire within material. MTF must be within +/- 0.5 lp/cm or +/- 15% of baseline.
21
How to perform CT Spacial Resolution Test
1. usually uses bar pattern test device direct measurement test tool or device with repeated pattern of holes, bars or lines (lp/cm) 2. image and evaluate for smallest line pari visible
22
CT Low Contrast Detectability/Low Contrast Resolution
Determines smallest object at a specified contrast level that can be seen, under regular reconstruction conditions. SC35 requires this test be done quarterly.
23
How to perform CT Low Contrast Resolution Test
1. uses phantom with various sized objects, with low density difference 10 HU and less from background 2. image and visually evaluate for presence of objects in image
24
CT Patient Support Movement Test
Determines accuracy of patient movement to ensure correct volume is scanned. SC35 requires this test done quarterly. Move table with "patient" phantom/weight and measure with ruler. Must be with +/- 1 mm
25
CT Laser Light Accuracy
Determine the accuracy of positioning light. SC35 requires this test be done semi-annually. Use thin wire of 1mm and ruler or vendor test tool. Tool/wore imaged based on laser and image evaluates. Axial scan localization light must be within +/- 2 mm, sagittal coronal and isocenter alignment should be within +/- 5 mm.
26
CT Patient Dose Test
Determines the CTDI 100 for head and body scans. SC35 requires this test be done semi-annually.
27
How to perform CT Patient Dose Test
1. dosimeter phantoms, dosimeter with dose probes (100mm pencil ionization chambers) at recommended scanning protocols 2. ionization chamber dose measured at centre of phantom and at 12, 3, 6 and 9 o'clock peripheral positions and in air (external to phantom) 3. repeated for head and body, different protocols 4. calculate CTDI 100. Must be within +/- 20% manufacturer specifications, should be within +/- 10 %