Week 6 (spectroscopic methods) Flashcards
(37 cards)
What are the different spectroscopic methods?
- untraviolet-visible spectroscopy (UV-VIS)
- fluorescence spectrophotometry
- atomic absorption spectroscopy (AAS)
- Nuclear Magnetic spectroscopy (AAS)
Explain ultraviolet visible spectroscopy (UV-VIS)
- All organic compounds absorb UV light, sometimes at very short wavelength
- UV energy is similar to the bonding energy in organic molecules
- The UV-Vis light is passed through a sample and the transmittance of light by a sample is measured
- An absorbance spectrum is obtained that shows the absorbance of a compound at different wavelengths. The amount of absorbance at any wavelength is due to the chemical structure of the molecule
Each wavelength of light has a particular energy associated with it. If that particular amount of energy is just right for making one of these energy jumps, then that wavelength will be absorbed - its energy will have been used in promoting an electron to higher energy (empty molecular orbitals)
The larger the energy jump, the lower the wavelength of the light absorbed.
UV-Vis can be used in a qualitative manner, to identify functional groups or confirm the identity of a compound by matching the absorbance spectrum. It can also be used in a quantitative manner, as concentration of the analyte is related to the absorbance using Beer’s Law
ΔE= E2-E1=hv
E= energy absorbed or emitted
h= Planck’s constant
v = frequency of radiation
What is Beers law?
- Light absorption is proportional to concentration of absorbing species
What is Lamberts law?
Light absorption is proportional to path length
What is the Beer-Lambert law?
A= log10(Io/I)= εcl
A= absorbency
Io= Incidence light intensity
I= transmitted light intensity
εh= molar absorbance coefficient at wavelength h
c= concentration of sample
l= path length (cuvette size)
Units are often: dm3mol-1cm<span>-1</span>
Explain the components of the Beer-Lambert law equation: A=εcl
A=εcl
where A is absorbance, ε is the molar extinction coefficient (which depends on the nature of the chemical and the wavelength of the light used), l is the length of the path light must travel in the solution in centimetres, and c is the concentration of a given solution
What are some clauses for deviation from the Beer-Lambert law?
Sample
- contamination (with material that’s absorbs in the same region of the UV spectrum)
- precipitation
- degradation (photolysis)
- fluorescence
- tautomerisation
- pH effects
- temperature
What are some other causes for deviation from the Beer-Lambert law?
- stray light
- non- monochromatic light source
- mismatched cells
- sensitivity A< 0.002
- solvent absorption
What must happen for each sample under the measurement conditions to be used over an appropriate concentration range?
Must establish the validity of the Beer-Lambert law
Calibration curves:
- use at least 5 standard solutions spanning the working concentration range
- Measure in duplicate in a matched pair of cells against the solvent as reference
Multicomponent systems:
- the absorption spectra often overlap
- if the components obey the Beer-Lambert law AND the law of additivity of absorbance applies then the observed spectrum is simply the sum of the components
- simultaneous equations (one per component)
What do multicomponent systems need?
- accurate absorptivity values
- non- overlapping hmax regions for components
- the errors are very great for similar components
- make derivatives for colourmetric analysis?
Explain fluorescence spectrophotometry
It involves using a beam of light, usually ultraviolet light, that excites the electrons in molecules of certain compounds and causes them to emit light; typically, but not necessarily, visible light.
- very sensitive
- better than absorption spectrophotometry (where Io/I is difficult at low concentration)
- measured against a “dark” background (Incident light is at a different wavelength to what is measured- result light)
- selective - fluorescent drugs or biomolecules and their metabolites May be analysed more readily than by conventions, spectroscopy
Using the diagram how does it work?
What are the applications of fluorescence spectrophotometry?
There is a light source and a monochromator (something that allows selection for the wavelength of light). The light goes in and it meets molecules in the cuvette and excites some of them, flurescence energy comes out in all directions. Another monochromater and detector is organised perpendicular to the initial light source (avoid the light passed directly through arriving on the detector)
- enzymes assays and kinetic analysis
- protein structure
- membrane structure
- fluorescence bleaching recovery
- energy transfer studies - fluorescence depolarisation
- microspectrofluorimetry
- fluorescence immunoassay
- Fluorescence activated cell sorting
- multicomponent analysis by synchronous luminescence spectrophotometry
Fluorescence spectrophotometry: What is signal intensity affected by?
- generally more sensitive to environment than absorbance measurements
- the signal intensity may be affected by
- pH (any ionisable groups)
- temperature (T^, increasedcollisions, quenching - use a thermostat for precise work)
- quenching (formation of complex between the sample and another species-photons absorbed by molecules)
- interfering substances
- often limiting in the analysis of biological samples; can be reduced by pre treatment of the sample, use of pure solvents, clean glassware)
- solvent
- interference from Rayleigh and Raman scattering
Explain the Jablonski diagram
- In molecular spectroscopy, a Jablonski diagram is a diagram that illustrates the electronic states of a molecule and the transitions between them. The states are arranged vertically by energy and grouped horizontally by spin multiplicity
- Electronic ground state with several vibrational energy levels to create a vibrational energy ladder and electronic excited state also with several vibrational energy levels
- Radiative transitions involve the absorption of a photon, if the transition occurs to a higher energy level, or the emission of a photon, for a transition to a lower level.
- For flourescent comounds, the emission wavelengths is longer than the excitaion wavelength
Fluorescence spectrophotometry: Quantitative applications
In dilute solution fluorescence intensity If = KQIoebc
If proportional to c
- k= instrumental constant
- Q= quantum efficiency of fluorescence
- Io= incident light intensity
- eh= molar absorbance coefficient at wavelength h
- c= concentration of sample
- b= path length
Explain Fluorescence spectrophotometry
- If an atom or molecule first absorbs energy – for instance a photon -, this is called excitation. Very shortly (in the order of nanoseconds) after excitation it emits a photon of a longer wavelength. We call that fluorescence
- In fluorescence spectroscopy, a beam with a wavelength varying between 180 and ∼800 nm passes through a solution in a cuvette. We then measure – from an angle - the light that is emitted by the sample.
- In fluorescence spectrometry both an excitation spectrum (the light that is absorbed by the sample) and/or an emission spectrum (the light emitted by the sample) can be measured. The concentration of the analyte is directly proportional with the intensity of the emission.
What must the total absorbance of the system (εcl) not exceed?
0.005 absorbance units
What may happen at high drug concentrations?
Ground-state molecules may absorb the fluorescence emitted by excited molecules
>negative deviations from linearity
What must the range of linearity of the calibration curve (If vs c) use?
at least 5 standard solutions
Unlike absorption spectrophotometry:
Spectrofluorimetry does not give an absolute scale of values.
> MUST USE A REFERENCE STANDARD for quantitative work
What is fluorescence derivatisation?
Some weakly fluorescent or non- fluorescent molecules may be derivatised with a fluorescent reagent e.g:
- Dansyl chloride Derivatives primary and secondary amines and phenols
- Fluorescamine Derivatives primary aliphatic amines including peptides and proteins
- Isatonic anhydride Derivatives primary and secondary amines and alcohols
- o-Phthaldehyde Derivatives primary amines
What are some considerations of fluorescence derivatisation?
- fluorescent contaminants
- extent derivatisation
- reliability of the derivatisation reaction
