0-RT103Ch1+RT207Ch1,2,3 Flashcards
Cathode
Cathode = Filament
a. dual-focus
b. Focusing cup
IS
Intensifying Screens: PRODUCE LIGHT in response to x-rays, greatly REDUCING patient DOSE
mAs in Inverse Square Law
mAs affects density (darkness of image) and is used to compensate for changes in SID according the the following:
mAs^1 / mAs^2 = (SID^1)squared / (SID^2)squared
5 Characteristics of Radiation
- Travels in STRAIGHT lines at LIGHT SPEED
- AFFECTS photographic EMULSIONS
- Affects biologic tissues (IONIZES ATOMS)
- CanNOT be FOCUSED or REFRACTED
- UNDETECTABLE human senses
Exposure Indicators
Only method determining if an image has been over- or underexposed
4 Settings Found on X-Ray Tube Control Panel
- mA
- kVp Major
- kVp Minor
- Time
Common IR Sizes (Film + Cassette)
8x10 in (20x25 cm) 10x12 in (25x30 cm) 11x14 in (28x35 cm) 7x17 in (18x43 cm) 14x14 in (35x35 cm) 14x17 in (35x43 cm)
14 RT Duties in Fluoro Exam
- Obtain PATIENT’S HISTORY including appropriate preprocedural preparation (eg diet, bowel cleansing)
- Complete necessary preprocedural PAPERWORK (eg required consent, education forms)
- ASSIST PATIENT in disrobing and gowning
- EXPLAIN procedure to patient
- Produce required preliminary (SCOUT) IMAGES
- SET CONTROL PANEL correctly for fluoroscopy and spot-film radiography
- POSITION patient for start of procedure
- PREPARE equipment for fluoroscopy
- INPUT patient DATA into computer for digital imaging (if applicable)
- Load spot-film device (if applicable)
- Prepare appropriate CONTRAST AGENT(S)
- ASSIST radiologist
- Take follow-up radiographs
- Provide postprocedural care and instructions to patient
George Eastman
1914: Invented first X-ray film base
Image Detail
Refers to SHARPNESS of Image:
a. High Detail: edges and lines making up image are crisp and precise
b. Low Detail: lines/edges less distinctive and appear somewhat blurred or “out of focus”
CR
Computed radiography: Uses an imaging plate that appears similar to a cassette:
- contains a PHOTOSTIMULABLE PHOSPHOR rather than film
- image is READ BY LASER and is displayed on a monitor
3 Steps to Warm Up X-Ray Tube
- Turn on Power On Wall
- Turn on Power on X-Ray Tube Control Panel
- Set all settings on Control Panel to Lowest
SID
Source-to-Image Receptor Distance: Distance from the x-ray tube (Source) to the IR.
- Affects x-ray beam intensity and beam divergence (field size)
4 Avenues for Radiography Education
- On-the-job training evolved into hospital-based education.
- 1950’s: radiation therapy education became separate from radiography
- Community colleges
- Four-year universities
4 Basic Requirements of X-Ray Production
- Vacuum (x-ray tube)
- Electron source (filament)
- Target for electrons (anode)
- High potential difference (voltage) between the electron source and the target (opposite charges attract)
Film Storage
- Stored in an area protected from radiation and chemical processing fumes (reduces risk of fog)
- Boxes stored upright on end
- Oldest film used first (before expiration date)
- Proper handling reduces artifacts:
a. clean, dry hands
b. clean work surfaces
c. No bending
Crooke’s Tube
1870’s-1880’s: First device that produced X-rays
kV
Kilovoltage: Measure of the POTENTIAL DIFFERENCE across the x-ray tube.
- Determines SPEED OF ELECTRONS in the electron stream
- Provides KINETIC ENERGY each electron has when it collides with the target, which determines the x-ray photon energy, expressed by the wavelength of the photons
Michael Idvorsky Pupin
Demonstrated the radiographic use of FLUORESCENT SCREENS (predecessor to intensifying screens)
Grids and Buckeys
Devices used to reduce the effect of scatter radiation on image quality; PREVENTS SCATTER from reaching the IR.
Image Intensifier
Allows image to be viewed in ambient light rather than in the dark (original images were very dim)
Cassettes
Film holder with both sides lined by intensifying screens
William Coolidge
Designed the first “HOT CATHODE” X-ray tube which became the prototype for modern X-ray tubes
Optical Density
Determined as the OVERALL BLACKNESS of an image;
In film/screen, density is controlled by mAs (Greater mAs = Darker image)
BUT
In digital imaging, the computer processing of the image doesn’t allow image to become darker from using more mAs; Exposure indicator value is the only way to determine overexposure
Overview of Radiographic Procedure
- X-ray photons pass through the patient in a variety of interactions based on the tissue types
- The x-ray photons that exit the patient strike the IR to create the latent image
- The manifest image is obtained depending the the IR type: Viewed on film or on a monitor
Thomas Edison
- Developed FIRST FLUOROSCOPE.
- Developed many of the FLUORESCENT MATERIALS used in today’s imaging
EM Energy
Electromagnetic Energy: X-Rays have electrical and magnetic properties, traveling 90 degrees from eachother in a straight line at the speed of light.
Primary X-Ray Beam and 4 Surrounding Parts
- Primary X-ray beam
- Focal spot
- Radiation field
- Central ray
- Collimator
Anode
Anode = Target
4 Main Duties of RT
- POSITION the patient’s ANATOMY to be imaged over the image receptor (IR)
- ALIGNS the x-ray TUBE (beam or central ray)
- SETS exposure FACTORS
- ACTIVATES the exposure SWITCH
Transformer
PROVIDES the HIGH VOLTAGE necessary to produce x-rays; Connected to the x-ray tube via cables
Fluoroscent Equipment
- Provides dynamic anatomic images (i.e. Images in motion).
- Allows radiologist to view and record images at the same time.
Photon
Smallest unit of EM energy
Image Contrast
Defined as the DIFFERENCE in the optical DENSITY of adjacent STRUCTURES within the image (primarily controlled by kVp).
3 Basic Steps of Film Processing
- Film is removed from the cassette and is placed in the processor in a darkroom.
- Cassette is reloaded with unexposed film from the film bin usually located under the counter
- Passboxes may be used to transfer cassettes to/from darkroom
SHORTER wavelengths = ___ energy and ___ penetration of tissues
SHORTER wavelengths = MORE energy and GREATER penetration of tissues
mAs
Product of mA and exposure time (in seconds):
mA x time (s) = mAs
An Increase in kVp = _______ Wavelength X-Rays
An Increase in kVp = SHORTER Wavelength X-Rays
Effects of SID and OID on Detail
Increase in SID = Increased Detail
Decrease in OID = Decreased Detail
Charles Daly
Thomas Edison’s ASSISTANT;
1904: FIRST DEATH from X-ray exposure in US (which caused Edison to cease his X-ray research)
2 Basic Types of Filmless Systems
- Computed radiography (CR)
2. Digital radiography (DR)
mA Settings for Most Machines
Most machines have mA settings from 50 to 500 (in increments of 50 to 100)
- Others may have higher settings up to 1500 mA
- mA setting also selects the large or small filament (focal spot size) where higher mA selects larger focal spot sizes.
Shape Distortion
Caused by UNEQUAL MAGNIFICATION of the parts in the subject:
a. Part PARALLEL with the IR = LESS distortion
b. Central ray PERPENDICULAR to the part = LESS distortion
mA
Milliamperage: Indication of the NUMBER of x-ray PHOTONS produced per SECOND.
Higher kVp = ___ contrast (___shades of gray)
Higher kVp = LOWER contrast (MORE SHADES of gray)
4 Factors of Image Quality
- Density
- Contrast
- Detail
- Distortion
Visibility of image is determined by overall blackness and differences between the blackness (density and contrast)
Greater mAs = _____ x-ray photons = _____ image
Greater mAs = MORE x-ray photons = DARKER image
February 1896
FIRST documented MEDICAL USE of X-ray in US (boy’s fractured wrist) at Dartmouth College
Wilhelm Roentgen
- Nov 8, 1895: DISCOVERED X-RAYS produced by CROOKE’S TUBE.
- Produced the FIRST ANATOMIC RADIOGRAPH (his wife’s hand)
Lower kVp = ___contrast (___ shades of gray)
Lower kVp = HIGHER contrast (FEWER shades of gray)
X-Ray Tube Support
Provides support and mobility for the tube. 2 Types: 1. Ceiling mount (photo) 2. Floor stand Electronic locks keep tube in place.
Film
Composed of polyester base coated on both sides with emulsion; made to be sensitive to light emitted by IS
Technique Charts
Provide recommended prime exposure factors for various exams and patient sizes.
Some machines have “anatomic programs” which sets recommended techniques based on part and size
Upright Image Receptor Units
Device that holds the IR and/or a bucky (grid) in a vertical position
H.C. Snook
Discovered INTERRUPTERLESS TRANSFORMERS that provided the necessary electricity for X-ray production
Higher kVp = ___ penetrating x-ray beam = ___ image
Higher kVp = MORE penetrating x-ray = DARKER image
Scatter Radiation
Created when a PORTION of an x-ray photon’s ENERGY is ABSORBED;
Primary SOURCE of occupational radiation EXPOSURE
Size Distortion is _____
Size Distortion is MAGNIFICATION, affected by SID and OID:
Increased SID or Decreased OID = Less Magnification
First X-Ray Technicians
Assistants to/followed by/trained by Physicians who developed first x-ray procedures to demonstrate anatomic structures
Sine Waves
REPEATING SINUSOIDAL WAVEFORMS created by changes in the EM field
Inverse Square Law
States that the intensity of the radiation is inversely proportional to the square of the distance.
I^1/I^2 = (D^2)squared / (D^1)squared where: I^1 = original beam intensity I^2 = new beam intensity D^1 = original SID D^2 = new SID
Quanta
BUNDLES of photons
Distortion
Refers to the VARIATION of the SIZE or SHAPE of the IMAGE in COMPARISON the to OBJECT it represents
DR
Digital Radiography: Has no cassettes and doesn’t require a separate processor/reader.
- After exposure, the image appears on the monitor
SID
Source-to-Image Distance: Distance from Anode (Source) to the IR, affecting MAGNIFICATION, recorded DETAIL and patient DOSE.
rem = ___ x ___
rem = rad x WF
4 Types of Distortion
- Shape distortion
- Elongation
- Foreshortening
- Magnification
5 Ways of Controlling Voluntary Motion
- Giving clear instructions
- Providing patient comfort
- Adjusting support devices
- Applying immobilization
- Decreasing exposure time
Density
Degree of blackening (brightness in digital imaging); Controlled by mA, exposure time and mAs.
6 Factors Affecting Recorded Detail
- Geometry (shape of part)
- Film
- Distance
- Screen
- Focal spot size
- Motion (main cause of lack of detail)
CR
Computed Radiography: Uses a Phosphor IMAGING PLATE (IP) as its IR.
- After exposure, IP is read by digital reader (computed)
- Image is displayed on monitor
NOTE: Optimum kVp isn’t essential to image quality but slightly higher kVp yields better images
Lat
Lateral
Ethics
Profession’s moral responsibility and the science of appropriate conduct towards others; Current code developed by ARRT for US
3 Pieces of Info that RT Must Verify
- Spell last name
- DOB
- Ask procedure that patient is there for
7 Causes of Voluntary Motion
- Nervousness
- Discomfort
- Excitability
- Mental illness
- Fear
- Age
- Breathing
Coulombs/kg (C/kg)
SI unit of exposure
2 Systems of Measurement
- Traditional (British)
2. Systeme Internationale (SI)
Basic Patient Care
RT is responsible during imaging procedure.
CRITICAL for RT to obtain patient’s clinical history:
- verify correct procedure ordered
- observe conditions of abnormalities to relay to radiologist
3 Effects of SID Changes
- Magnification: longer SID reduces magnification
- Recorded Detail: longer SID increases recorded detail
- Patient Dose: longer SID decreases patient dose
Rule #3 of Radiography
- Must place the part as close to the IR as possible.
4 Primary Exposure Factors
- Exposure time (T) in seconds (s)
- Milliamperage (mA)
- Kilovoltage Peak (kVp)
- Source-to-Image Distance (SID)
Image Display
Radiographs usually viewed in anatomical position, viewed from perspective of X-ray tube with distal ends towards ceiling.
EXCEPTIONS: hands, feet, wrist, toes.
3 General IR Positions
- Longitudinal aka Length-wise (“portrait”)
- Horizontal aka Cross-wise (“landscape”)
- Corner-to-Corner aka Diagonal (“diamond”)
NOTES:
a. Position name based on IR relation to long axis of body
b. Longitudinal placement most often used
Magnification
Type of size distortion, present in all images; Controlled by OID and SID
5 Factors Causing Distortion
- Alignment
- Central ray
- Anatomical part
- IR
- Angulation
Distortion
Misrepresentation of the size/shape of structure; Caused by alignment, central ray, anatomical part, IR and angulation.
Dosimeters
Devices for monitoring personnel radiation exposure:
Worn at collar with label facing out
3 Factors Affecting Density
- mA
- Exposure time (seconds)
- Milliampere-seconds (mAs)
8 Rules for Specific Marker Placement
- For AP and PA projections that include R and L sides of body (eg head, spine, chest, ABD, pelvis, R marker typically used.
- For LAT projections of head and trunk (eg head, spine, chest, ABD, pelvis), always mark side closest to IR.
- For OBL projections that include R and L sides of body (eg spine, chest, ABD), the side down, or nearest IR, is typically marked. For a Right Posterior Oblique (RPO) position, mark R side.
- For Limb Projections, use appropriate R or L. Marker must be placed within edge of collimated beam.
- For limb projections done with 2 images on same IR, only 1 of the projections needs mark.
- For limb projections where R and L sides are imaged side by side on 1 IR (eg R and L, AP knees), R and L markers required to identify each side clearly.
- For AP, PA or OBL chest projections, marker is placed on upper-outer corner so that thoracic anatomy isn’t obscured.
- For decubitus positions of chest and ABD, R or L marker should always be placed on the side up (opposite the side laid on) and away from anatomy of interest.
Nonstochastic Effects
Occur only after a certain amount of exposure has been received with severity being dose-dependent.
OBL
Oblique: halfway between AP (or PA) and Lat
Density Comparisons
- Insufficient density (soft tissue and bone look same)
- Optimum density
- Excessive density (too dark/no soft tissue)