Belmont Rapid Infuser Flashcards

(7 cards)

1
Q

Belmont Rapid Infuser

What is the Belmont and what is it used for?

Belmont Rapid Infuser

A

The Belmont Rapid Infuser RI-2 is a high-performance infusion device built for emergency or high-volume fluid administration scenarios.

It’s very useful during:
* Trauma resuscitation (massive transfusion)
* Major surgeries (cardiothoracic, orthopedic, transplant)
* Hypothermic patients
* Military medicine and field hospitals

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
2
Q

Belmont Rapid Infuser

Core Capabilities

A

The Belmont’s core capabilities are:
* Delivering fluids and blood from 2.5 to 1000 mL/min
* Heats fluids to physiologic temperature
* Incorporates automated air detection, pressure control, and safety cutoffs.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
3
Q

Belmont Rapid Infuser

Engineering Overview

A

Key components:
* Roller Pump: Reliable, consistent fluid delivery
* IR Probes (Input/Output): Ensure correct temperature management
* Fluid Out & Air Detectors: Stop pump if out of fluid or air is detected
* Valve Wand: Switches flow path instantly to ensure patient safety
* Touch Display: Intuitive interface for setup and troubleshooting
* Battery Backup: Maintains pumping (no heating) for 30 minutes.

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
4
Q

Belmont Rapid Infuser

Warning and Contraindicated Uses

A

Contraindicated uses:
* Platelets, cryoprecipitate, granulocytes
* Drugs or unprocessed blood
* Any calcium or dextrose-based solutions mixed with blood

Warnings:
* Never reuse disposable sets
* Do not leave unattended during use
* Priming with blood is prohibited - use only compatible fluids
* Battery mode disables heating - suitable only for short-duration, low-flow use.
* High temperatures may damage RBCs - must be discarded if overheated

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
5
Q

Belmont Rapid Infuser

Setup & Priming

How-to video

A
  1. Inspect components, secure unit (optional IV pole)
  2. Install the sterile, pre-assembled disposable set
  3. Use only compatible fluids (e.g., NS, anticoagulated blood)

Prime:
* Main system: 100 mL at 500 mL/min
* Patient line: 50-200 mL/min manually

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
6
Q

Belmont Rapid Infuser

Infusion Operation

A
  1. Start infusion at 10 mL/min
  2. Adjust rate as needed (up to 1000 mL/min)
  3. Use bolus mode for fixed-volume infusions
  4. Automatic air purging after every 500 mL delivered
How well did you know this?
1
Not at all
2
3
4
5
Perfectly
7
Q

Belmont Rapid Infuser

Post-Use

A
  • STOP pump, clamp lines, and dispose of single-use set
  • Clean and disinfect system per infection control protocol
  • Perform periodic operational and visual checks