Chapter 13 Flashcards
(213 cards)
exposure factor
the factors that influence and determine the quantity and quality of x-radiation to which the patient is exposed
four prime exposure factors
kVP, mA, exposure time, SID
kVP and mAs
the most important factors principally responsible for x-ray quality and quantity. focal spot size, distance, and filtration are secondary factors that may require manipulation for particular examinations
kVP
affects quality and therefore, beam penetrability,with increasing kVP, more x-rays are emitted, and they have higher energy and greater penetrability. but because they have higher energy, they also interact more by Compton effect and produce more scatter radiation, which results in reduced image contrast.
the kVP selected helps to determines the number of x-rays in the image forming beam, and hence the resulting average optical density (OD).
finally, and most importantly, the kVP controls the scale of contrast on the finished radiograph because as kVp increases, less differential absorption occurs, therefore, high kVP results in reduced image contrast.
Milliamperes
the mA selected determines the number of x-rays produced and therefore the radiation quantity. the unit of electric current is the ampere (A) one ampere is equal to 1 coulomb (c) of electrostatic charge flowing each second in a conducter.
fallng load generator
on an x-ray imaging system in which only mAs can be selected, exposure factors are adjusted automatically to the highest mA at the shortest exposure time allowed by the high voltage generator
Distance
affects exposure of the image receptor according to the inverse square law, the SID largely determines the intensity of the x-ray beam at the image receptor.
direct square law
is derived from the inverse square law, it allows a radiologic technologists to calculate the required change in mAs after a change in SID to maintain constant OD.
focal spot size
for general imaging, the large focal spot is used. this ensures that sufficient mAs can be used to image thick or dense body parts. the large focal spot also provides for a shorter exposure time, which minimizes motion blur.
one difference between large and small focal spots is the capacity to produce x-rays. many more x-rays can be produced with the large focal spot because anode heat capacity is higher. with small focal spot, electron interaction occurs over a much smaller area of the anode, and the resulting heat limits the capacity of x-ray production.
a small focal spot is reserved for fine-detail radiography, in which the quantity of x-rays is relatively low. small focal spots are always used for magnification radiography. these are normally used during extremity radiography and in examination of other thin body parts in which higher x-ray quantity is not necessary.
filtration
three types of x-ray filtration are used: inherent, added, and compensating.
inherent filtration
all x-ray beams are affected by the inherent filtration properties of the glass or metal envelope of the x-ray tube. the value of the inherent filtration is approximately 0.5 mm the required total filtration of 2.5 mm is needed.
compensating filters
are shapes of aluminum mounted onto a transparent panel that slides in grooves beneath the collimator. these filters balance the intensity of the x-ray beam so as to deliver a more uniform exposure to the image receptor. they may be shaped like a wedge for examination of the spine or like a trough for chest examination.
as added filtration is increased, the result is increased x-ray beam quality and penetrability. the result on the image is the same as that for increased kVP, that is, more scatter radiation and reduced image contrast.
three basic types of high voltage generators are available
single phase, three phase, and high frequency
high wave rectified generator
has 100% voltage ripple. during exposure with a half wave rectified generator, x-rays are produced and emitted only half the time. during each negative half cycle, no x-rays are emitted.
full wave rectification
identical to half wave rectification except there is no dead time. during exposure, x-rays are emitted continually as pulses. consequently, the required exposure time for full wave rectification is only half that for half wave rectification.
three phase power
comes in two principal forms: 6 pulse or 12 pulse. the difference is determined by the manner in which the high voltage step up transformer is engineered.
the difference between the two forms is minor but does cause a detectable change in x-ray quantity and quality. three phase power is more efficient than single phase power. more x-rays are produced for a given mAs setting, and the average energy of those x-rays is higher. the x-radiation emitted is nearly constant rather than pulsed.
patient factors
such as anatomical thickness and body compositon
image quality factors
such as OD, contrast, detail, and distortion
exposure technique factors
such as kVP, milliamperage, exposure time, and SID, as well as grids, screens, focal spot size, and filtration.
body habitus
sthenic – meaning “strong, active” patients are average.
hyposthenic–are thin but healthy appearing; these patients require less radiographic technique.
hypersthenic—are big in frame, and usually overweight.
asthenic are small, frail, sometimes emaciated, and often elderly.
calipers
are used to measure the thickness of the anatomy that is being irradiated
composition
when only soft tissue is being imaged, low kVp and high mAs are used. with an extremity, which consists of soft tissue and bone, low kVp is used because the body part is thin.
when imaging the chest, the radiologic technologist takes advantage of the high subject contrast. lung tissue has very low mass density, bony structures have high mass density, andd the mediastinal structures have intermediate mass density. consequently, high kVp and low mAs can be used to good advantage. this results in an image with satisfactory contrast and low patient radiation dose.
radiolucent
attenuates few x-rays and appears black on the radiograph
radiopaque
tissue absorbs x-rays and appears white on the radiograph