Chapter 31 (Module 2) Flashcards
Quality assurance includes each of the following EXCEPT
a. equipment maintenance.
b. radiology reports accuracy.
c. diagnosis of images.
d. scheduling.
Diagnosis of images
The largest hospital accreditation agency is the
a. MSQA.
b. The Joint Commission.
c. Det Norske Veritas (DVN).
d. FDA.
The Joint Commission
The Joint Commission approval is linked to ____ reimbursements.
a. federal
b. state
c. insurance company
d. all of the above
All of the above
Input for the purchase of new radiographic equipment should include the
a. radiologists.
b. radiographers.
c. administrators.
d. all of the above.
All of the above
Installation and acceptance testing of equipment is the responsibility of the
a. radiation physicist.
b. radiology supervisor.
c. radiology equipment vendor.
d. biomedical engineering department.
Radiology Equipment Vendor
Routine quality management procedures should be performed by the
a. radiation physicist.
b. qualified radiographer.
c. chief radiologist.
d. biomedical engineering staff.
Qualified radiographer
Which quality control should be done on a regular, ongoing basis?
a. collimator accuracy
b. focal spot size estimation
c. image repeat analysis
d. beam HVL
Image repeat analysis
Each of the following monitoring tests should be done on a semiannual basis EXCEPT
a. processor sensitometry.
b. focal spot size estimation.
c. collimator accuracy.
d. kilovoltage accuracy.
Processor sensitometry
Test tools for focal spot size estimation include each of the following EXCEPT a
a. pinhole camera.
b. star test pattern.
c. line pair resolution tool.
d. motorized synchronous top.
Motorized synchronous top
Focal spots measuring 0.3 mm or less should be measured with a
a. star test pattern.
b. pinhole camera.
c. spinning top.
d. dosimeter.
Star test pattern
Focal spots measuring 0.3 mm or more should be measured with a
a. star test pattern.
b. pinhole camera.
c. spinning top.
d. SMPTE pattern.
Pinhole camera
Which type of equipment is used to measure half-value layers?
a. wire mesh test tool
b. goniometer
c. dosimetry equipment
d. DUKE chest phantom
DUKE chest phantom
When testing collimator accuracy, an SID error of ____ is allowed.
a. 20 percent
b. 10 percent
c. 2 percent
d. 0 percent
2 percent
The centering mark should be within ____ of the light field central ray.
a. 0 percent
b. 1 percent
c. 11 percent
d. 100 percent
1 percent
Distance indicators are checked with a
a. chalk line.
b. cell phone.
c. laser.
d. tape measure.
Tape measure
A generator must be recalibrated when the kVp drifts beyond ____ kVp of the labeled setting.
a. 2
b. 5
c. 10
d. 12
5
The accuracy of single-phase generator time settings can be measured with a
a. sensitometer.
b. densitometer.
c. computerize dosimeter.
d. penetrometer.
Computerize dosimeter
A thorough reject analysis program would include data collection on images produced
a. by each technologist.
b. in each radiographic room.
c. by service engineers.
d. all of the above.
All of the above
Exposure time settings should be maintained within ____ of the labeled setting.
a. +/- 5 percent
b. +/- 10 percent
c. +/- 15 percent
d. 0 percent
+/- 5 percent
The source-to-skin distance limit for a stationary fluoroscopic unit is ____ inches.
a. 5
b. 12
c. 15
d. 25
15
Tomographic system uniformity and completeness of motion are tested with a
a. pinhole camera.
b. wire mesh.
c. lead mask with a pinhole.
d. moving pendulum angle indicator.
Lead mask with a pinhole
Tomographic systems should be tested for each of the following EXCEPT
a. automatic brightness control.
b. completeness of motion.
c. section depth indicator accuracy.
d. section thickness accuracy.
Automatic brightness control
Most AECs utilize a ____ ion chamber combination.
a. single
b. two
c. three
d. four
Three
Computerized storage and transmission systems for digitized images of all types are called
a. DSR.
b. MRI.
c. CT.
d. PACS.
PACS