Equipment Flashcards
(91 cards)
22.2 Which is least likely to cause inaccuracies in pulse oximetry
a) Anaemia
b) Vasoconstriction
c) AF
d) Methaemoglobin
e) Carboxyhaemoglobin
a) Anaemia
No effect
- Fetal haemoglobin (HbF)
- SulphHb
- Bilirubin (absorption peaks are 460, 560 and 600 nm)
- dark skin
Falsely low reading
1. Methaemoglobin (MetHb). The presence of MetHb will prevent the oximeter from working accurately and the readings will tend towards 85%, regardless of the true saturation.
2. Methylene blue. When methylene blue is used in surgery (e.g. parathyroidectomy or to treat methaemoglobinaemia), a short-lived reduction in saturation estimations is seen. Readings may fall by 65% at a concentration of 2-5 mg/kg for between 10 and 60 minutes.
3. Indocyanine green. Use of this dye (e.g. in cardiac output studies) may cause a transient reduction in recorded saturations.
4. A reduction in peripheral pulsatile blood flow produced by peripheral vasoconstriction results in an inadequate signal for analysis.
5. Venous congestion, which may be caused by tricuspid regurgitation, high airway pressures and the Valsalva manoeuvre, may produce venous pulsations which can produce low readings.
6. Venous congestion of the limb may affect readings, as can a badly positioned probe.
7. External fluorescent light in the operating theatre may cause the oximeter to be inaccurate, and the signal may be interrupted by surgical diathermy. Shivering may cause difficulties in picking up an adequate signal.
8. Nail varnish may cause falsely low readings.
Falsely high reading
1. Carboxyhaemoglobin (CoHb). CoHb (haemoglobin combined with carbon monoxide) is registered as 90% oxygenated haemoglobin and 10% desaturated haemoglobin - therefore the oximeter will overestimate the saturation.
Calibration
- Oximeters are calibrated during manufacture and automatically check their internal circuits when they are turned on.
- They are accurate in the range of oxygen saturations of 70% to 100% (+/-2%), but are less accurate under 70%. Below the saturation of 70%, readings are extrapolated.
- The data for calibration came from human volunteer studies, hence it was unethical to allow the saturations to fall below 70%. Due to the shape of the oxyhaemoglobin curve, the saturation starts to fall rapidly at 90%.
Limitations
- The oximeter averages its readings every 10-20 seconds. Hence, they cannot detect acute desaturation. The finger probe has a response time of approximately 60 seconds, whereas the ear probe has a response time of 10-15 seconds.
- The site of application should be checked at regular intervals, as pressure sores and burns have been reported.
- The pulse oximeter only provides information about oxygenation. It does not give any indication of the patient’s carbon dioxide elimination.
N22.2 The piece of airway equipment shown is a
a. bullard laryngoscope
b. CMAC video stylet
c. lightwand
d. flexible bougie
CMAC video stylet
see image for alternative equipment images
21.1 The equipment shown in the picture is a (airway device shown)
a) Arndt bronchial blocker
b) Cohen bronchial
blocker
c) Microlaryngoscopy tube
d) Hunsaker tube
e) Parker flex ETT
Hunsaker Mon-jet ventilation tube for microlarnygeal surgery
Description:
-Laser-safe
-fluoroplastic
-self-centring catheter
Uses:
-subglottic ventilation during microlaryngeal surgery
Components:
- proximal end for attaching to jet insufflation system
-proximal end allows passage of stylet to aid insertion
-Side port at proximal end for monitopring airway pressure and ETCO2
-Outer diameter 4.3mm for maintaining good surgical access
-Green basket to keep the centre port at its tip away from tracheal mucosa and avoiding potential damage from jet ventilation
21.1 A respiratory effect of high flow nasal oxygen therapy is
A. Reduced RR
B. Reduced MV
C. Increased work of breathing
A. Reduced RR
BJA HFNOT
It has been demonstrated that patients with acute hypoxaemic respiratory failure experience improved comfort and tolerance with HFNOT compared with humidified oxygen via a facemask, and traditional non-invasive ventilation masks. Subjective feelings of dyspnoea AND RESPRIATORY RATES are REDUCED as is airway dryness.
23.1 In order to minimise the risk of cardiac arrhythmia, surgical diathermy has been designed to operate with
A. High frequency
B. High amplitude
C. Low frequency
D. Low amplitude
E. Using EES
A. High frequency
22.2 Based on this ECG tracing, the mode in which this pacemaker is operating is
a) VAI with intermittent failure to capture
b) AAI with intermittent failure to sense
c) DDD
d) VVI with intermittent failure to capture
e) VVI with intermittent failure to sense
e) VVI with intermittent failure to sense
21.1 You are performing a regional block for analgesia following knee surgery. You have an ultrasound probe scanning the anterior mid-thigh. The muscle indicated by the arrow in the ultrasound image below is the
A: Sartorius
B: Vastus Medialis
C: Adductor Longus
D: Gracilis
E: Rectus femoris
A: Sartorius
20.2 You are performing a regional block for analgesia following knee surgery. You have an ultrasound probe scanning the anterior mid-thigh. The muscle indicated by the arrow in the ultrasound image below is the
A. biceps femoris
B. Sartorius
C. Gracillis
D. Adductor longus
E. Adductor magnus
Sartorius
repeat
22.2 The Pin Index System positions on a C size cylinder of medical oxygen are
a) 1,5
b) 2,5
c) 3,5
d) 1,6
e) these options are made up
b) 2,5
Air: 1, 5
Oxygen: 2, 5
N2O: 3,5
CO2: 2, 6
He: 2, 4
Cyclopropane 3, 6
Entonox 7
21.1 The advantage of the Mapleson E circuit in paediatric anaesthesia is due to its
A. Can use low gas flows
B. Feel compliance
C. Assess tidal volume
D. Can rapidly change levels of CPAP
E. Low resistance
low resistance
MAPLESON E
- Derived from the Ayre T-piece used in Mapleson D circuit and functions on the same principle as Mapleson D
- The primary difference is in the length of the tubing that is increased to be greater than the patient’s tidal volume
- For spontaneous ventilation, the expiratory limb is open to the atmosphere
- It has no valves so there is no resistance to airflow nor points for possible mechanical failure
- Rebreathing is dependent on the fresh gas flow, patients minute volume and capacity of the expiratory limb
- Its main use is in paediatric patients
20.1 This type of tracheal tube is best described as a (picture of airway device shown)
a) Mini tracheostomy tube
b) South facing RAE
c) Laser tube
d) Laryngectomy tube
e) Fenestrated tracheostomy tube
laryngectomy tube
Rusch Larygoflex Reinforced Laryngectomy tube -
22.2 A patient is anaesthetised from the awake state to a state of surgical anaesthesia with propofol or a volatile anaesthetic. As the depth of anaesthesia increases, the patient’s electroencephalogram (EEG) will show oscillations that are of
a. low frequency low amplitude
b. low frequency high amplitude
c. high frequency low amplitude
d. high frequency high amplitude
b. low frequency high amplitude
Changes in the electroencephalogram during anaesthesia and their physiological basis
https://academic.oup.com/bja/article/115/suppl_1/i27/234261
Figure 1 shows raw EEG waveforms during isoflurane anaesthesia.
During light anaesthesia:
-amplitude is shallow and frequency is high.
When a higher concentration is administered:
-amplitude deepens and EEG frequency slows.
During deep anaesthesia:
- a ‘burst and suppression’ pattern becomes apparent, characterized by extreme activity, represented by high-frequency, large-amplitude waves (bursts), alternating with flat traces (suppression).
- This pattern, excluding brain ischaemia or other factors, indicates that anaesthesia is too deep. Beyond this, flat traces become dominant and, eventually waveforms are no longer apparent.
During isoflurane, sevoflurane or propofol anaesthesia, this sequence of changes in pattern is almost identical.
The major difference in EEG between the volatile agents (isoflurane or sevoflurane) and propofol is apparent in power in the theta range.
During propofol anaesthesia, theta power remains low regardless of concentration, but during isoflurane or sevoflurane anaesthesia, it increases at surgical concentrations of anaesthesia.
22.1 A drug which does NOT increase the defibrillation threshold in a patient with an implanted cardioverter defibrillator is
a. Amiodarone
b. Atropine
c. B-blocker
d. Flecainide
e. Sotalol
e. Sotalol
Drugs that INCREASE defibrillation threshold:
+ Amiodarone (Chronic)
+ Atropine
+ lignocaine
+ Diltiazem
+ Flecainide
+ Verapamil
+ Venlafaxine
+ Anaesthetic agents.
Drugs that DECREASE defibrillation threshold:
- Sotalol
- Amiodarone (acute)
- Nifekalant
Drugs with No Change in DFT
= B- blocker
= Disopyramide
= Procainamide
= Propafenone
https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6304797/
23.1 During standard diagnostic nocturnal polysomnography for investigation of obstructive sleep apnoea, apnoea is defined as cessation of airflow for
A. 10 sec
B. 20 sec
C. 30 sec
D. 10 sec with 3% desat
E. 20 sec with 3 % desat
A
Apnea is defined as the cessation of airflow for ten or more seconds.
Hypopnea is defined as a recognizable, transient reduction, but not a complete cessation of, breathing for ten or more seconds.
Hypopnea requires a 4% fall in SpO2
https://www.ncbi.nlm.nih.gov/books/NBK441909/#:~:text=Obstructive%20Sleep%20Apnea%20(OSA)%2C,for%20ten%20or%20more%20seconds.
22.1 When using an endotracheal tube in an adult, the highest recommended cuff pressure to avoid mucosal ischaemia is
a. 10cmH2O
b. 20
c. 30
d. 40
e. 50
c. 30cmH2O
22.2 According to ANZCA PS54(A), an anaesthetic machine requiring electrical power must, in the event of mains power failure, be able to operate under battery backup power for a minimum of
a) 30 min
b) 60 min
c) 120 min
d) 240 min
a) 30 min
If the anaesthesia machine requires electrical power for normal operation, a backup power supply must be a part of the machine and permit normal operation for at least 30 minutes after a mains power supply failure. An alarm must be activated at the time of the mains failure and the state of the reserve power supply must be indicated while it is in use.
https://www.anzca.edu.au/getattachment/f05e02ec-2023-4c50-b57f-9549ea0c4183/PS54(A)-Position-statement-on-the-minimum-safety-requirements-for-anaesthesia-machines-and-workstations-for-clinical-practice-2021#page=
23.1 A level two check of the inhalational anaesthesia delivery device does NOT include checking the
A. Accurate delivery of volatile concentration from vaporiser
B. Connection of vaporiser and seating
C. Secure vaporiser cap
D. Adequate filling of vaporizers
E. Power to vaporiser
a) Accurate delivery of volatile concentration from vaporiser
PS31
Level two check should be performed at the start of each anaesthetic list.
4.2.3.2 Inhalational anaesthesia delivery devices (vapouriser)
4.2.3.2.1 Ensure electricity is connected to vapourisers that require it.
4.2.3.2.2 Check the anaesthetic liquid level is within marked limits.
4.2.3.2.3 Ensure all filling ports are sealed.
4.2.3.2.4 Check correct seating, locking and interlocking of detachable vapourisers or casettes.
4.2.3.2.5 Test for circuit leaks with a cassette installed or for each vapouriser in the “on” and “off” state.
4.2.3.3 Check for machine leaks upstream from the common gas outlet or breathing system, using a protocol appropriate for the anaesthesia delivery system.
20.1 Best resolution US probe for median nerve visualisation:
d) 5-10mHz
e) 6-13mHz
High frequency probe at 90 degrees to the skin
- to best visualise superficial structures have the probe at 90 degrees to the skin with a high frequency transducer
it is best to use high-frequency transducers (up to 10–15 MHz range) to image superficial structures (such as for stellate ganglion blocks) and low-frequency transducers (typically 2–5 MHz) for imaging the lumbar neuraxial structures that are deep in most adults.
22.1 A patient is anaesthetised from the awake state to a state of surgical anaesthesia with propofol or a volatile anaesthetic. As the depth of anaesthesia increases, the patient’s electroencephalogram (EEG) will show oscillations that are of
Dominant EEG frequency decreases, and amplitude increases with increasing concentrations of anaesthetic. End result is burst suppression
https://academic.oup.com/bja/article/115/suppl_ 1/i27/234261
Figure 1 shows raw EEG waveforms during isoflurane anaesthesia.
During light anaesthesia:
-amplitude is shallow and frequency is high.
When a higher concentration is administered:
-amplitude deepens and EEG frequency slows.
During deep anaesthesia:
- a ‘burst and suppression’ pattern becomes apparent, characterized by extreme activity, represented by high-frequency, large-amplitude waves (bursts), alternating with flat traces (suppression).
- This pattern, excluding brain ischaemia or other factors, indicates that anaesthesia is too deep. Beyond this, flat traces become dominant and, eventually waveforms are no longer apparent.
During isoflurane, sevoflurane or propofol anaesthesia, this sequence of changes in pattern is almost identical.
The major difference in EEG between the volatile agents (isoflurane or sevoflurane) and propofol is apparent in power in the theta range.
During propofol anaesthesia, theta power remains low regardless of concentration, but during isoflurane or sevoflurane anaesthesia, it increases at surgical concentrations of anaesthesia.
20.1 The radial artery pressure trace shown below is from a patient who has an intra-aortic balloon pump in situ. The device has been switched to 1:2 augmentation to assess the timing. The trace shows an augmented beat followed by an un-augmented beat. With respect to the augmentation, the trace shows
a. Correct timing
b. Early inflation
c. Late inflation
d. Early deflation
e. Late deflation
Bonus question
b. Early inflation
Waveform features:
> Diastolic augmentation (peak B) encroaches on the peak corresponding to unassisted systole (peak A) – the two peaks have merged and are barely distinguishable.
> There is no ‘sharp V’ or dicrotic notch between peaks A and B.
Early IAB inflation may result in:
> Premature closure of the aortic valve and possible aortic regurgitation, thus impairing left ventricular emptying. There may be an increase in LVEDV, LVEDP and PCWP.
> Increased left ventricular wall stress (afterload) and increased myocardial oxygen consumption will occur.
how to correct:
Delay the onset of IAB inflation, so that it inflates at the dicrotic notch resulting in a ‘sharp V’ (see the normal pressure waveform).
22.2 A 45-year-old man is ventilated in the intensive care unit and is in a critical state. His pulmonary artery wedge pressure is 26 mmHg, cardiac index is 1.7 L/minute/m2 and his PaO2/FiO2 ratio is 200 mmHg. A decision is made to place him on extracorporeal membrane oxygenation. The most appropriate mode is
a) VV ECMO
b) VA ECMO
c) Atrio-aorto ECMO
d) Ventriculo-atrial ECMO
b) VA ECMO
PaO2/FiO2 ratio
Mild: 200-300 = mortality 27%
Moderate = 100-200 mortality 32%
Severe < 100 = Mortality 45%
Cardiac Index
Normal: 2.5-4.2l/min
PAWP:
Normal 4-12mmHg
CI is low, PaO2/FiO2 ratio is mild, PAWP is high
22.1 An anaesthetised patient is ventilated and has standard monitoring plus a central venous line. As surgery commences, the line isolation monitor alarms, indicating a potential leakage current of greater than 5 mA from one of the power circuits in use. The most appropriate action is to
a) Ignore it
b) Disconnect non-essential
equipment one by one to identify fault
Line isolation monitor alarms when single fault in system. If the alarm is going off, the last piece of equipment plugged in is usually suspect and should be unplugged.
22.2 1 MAC of sevoflurane affects the sensory evoked potential signal for spinal surgery by
a) increased latency, increased conduction speed, increased amplitude
b) increased latency, decreased conduction speed, decreased amplitude
c) decrease latency, increased conduction speed, decreased amplitude
d) increased latency, increased conduction speed, decreased aptitude
Increased latency, decreased conduction speed, decreased amplitude