What is the minimum wet contact time for alcohol as an intermediate-level disinfectant?
Alcohol must remain in wet contact with the surface for a minimum of 5 minutes.
What system generates automated requests for patient care items?
Centralized patient order entry system.
Which body creates and revises standards for packing requirements for sterilizing water for irrigation?
USP-NF.
Which microorganism is used in the biological indicator for monitoring Ethylene Oxide?
Bacillus atrophaeus.
What is the reason behind double peel pouching?
It allows aseptic presentation of the sterile item without contamination.
What is the reason for a tamper-evident lock on a rigid container?
It indicates the container has not been opened or tampered with.
Where should textile packs be placed when loading the sterilizer with a mixed load of textile and metal goods?
On the top rack.
What will cause wet packs except for which wrapped items?
Wrapped items with missing CI.
Where should Bowie Dick tests be run?
In dynamic air removal steam sterilizers.
What is the sterilizer setting for a Bowie Dick test?
3 1/2 - 4 min exposure at 270-273 degrees F.
What is the purpose of instrument count sheets?
To provide a detailed list of tray contents.
Which information is not required on the SDS sheets?
Emergency phone number of physicians.
Which body system controls growth and maintains body functions and chemical messengers?
Endocrine system.
What is not recommended for device taping as an instrument identification method?
Place the tape on the ring of the instrument.
What is the correct way to measure hemostatic forceps?
From the distal tip to the bottom of the ring.
What is the correct way to measure tissue forceps?
From the proximal end to the tips.
What chemical process provides a corrosion-resistant finish on instruments?
Passivation.
Which cleaning machine is appropriate for removing debris from box locks and hinges?
Ultrasonic washer.
Before using packaging materials for sterilization, what temperature and humidity should they be kept at?
68-73 degrees F & 30% to 60% humidity for at least 2 hours.
During handwashing, how long should hands be lathered and scrubbed?
For at least 20 seconds.
When preparing and packaging instruments for steam sterilization, how must the instruments be positioned?
Completely disassembled and in open position.
What statement about sterile processing of instruments is INCORRECT?
Trays with lids should be opened.
What is designed to prevent surgical instrument tray wrappers from pipes, holes, and tears?
Corner protectors.
Where should the Bowie Dick test pack be placed in the sterilization chamber?
Above the drain on the bottom rack.