finals L10 Flashcards
(6 cards)
1
Q
beam alignment frequency
A
every 3 to 6 months
2
Q
beam alignment equipment
A
- Appropriate test object for the alignment of the x-ray beam and the light field. (Coins or other metal objects)
- A cylinder with an attenuator at its center, or another test object that can be used to assess the perpendicular incidence of the beam on the image receptor.
3
Q
Make an exposure using __ kVp and __ mAs, or the values recommended by the medical physicist. beam alignment
A
50kVp, 3mAs
4
Q
X-ray light field alignment:
Ensure that the deviation of the xray field from the light field does not exceed __ cm on any side for an SID of 100 cm or __% of any other SID.
A
±2cm, ±2%
5
Q
Centering:
Ensure that the coincidence of the collimator light beam crosshairs with the x-ray beam center does not exceed __cm at an SID of 100 cm
A
±1
6
Q
If a trend is observed when repeating this test, then the ___ should adjust the beam limiting device.
A
service engineer