PLANT TISSUE CULTURE LAB Flashcards
(24 cards)
Growth of protoplasts, cells, tissues, shoot tips, roots, anthers, embryos, flowers and meristems in a laboratory medium.
Tissue culture
any nutrient material prepared for the growth and cultivation plant tissue.
culture medium
process of making any material entirely free from living microorganism.
sterilization
condition that describes the freedom of plant materials, culture medium, confines of the culture vessel from contaminating microorganisms.
asepsis
subjecting of cultures to conditions favorable to the growth of the plant tissue
incubation
flow air currents in which air currents do not intermingle
laminar flow
(CEILINGS) Ceilings shall be at least ______ from the finished floor line.
2.4 m
(WALL) All wall tops and ledges shall slope at ______.
45 degree
(WALL) It should be coved to the floor-wall and at wall-to-wall junctions with a radius of __________.
50 mm - 60 mm
(WINDOWS) All windows shall be properly installed with _________.
16-mesh screen
(WINDOWS) Windowsill shall be at least _______ from the finish floor line.
1 m
(DOORS) Main doorways shall be at least _______ wide to allow easier movement of large equipment and facilities.
1.5 m
(DOORS) Door for staff, storage and chemical rooms should be at least ________.
0.8 m
(FLOOR) The floor slope should be ________ towards the drain.
2% - 4%
(MEDIA PREPERATION AND STERILIZATION AREA) The area shall be provided with working tables. Tables should be up to ______ wide if
it is to be access from both sides, and should not exceed _______ if it is accessible only from one side and the height should be ________.
1.2 m , 600 mm , 850 mm – 900 mm
(GROWTH ROOM) Ceilings shall be insulated with a minimum R-value of ________.
2.64
(LIGHTING) Artificial lighting should be provided for the production of photo-period sensitive plants. For the workers, _______ should be provided.
200 lux
(GROWTH ROOM) The width of the shelves should be _________ if accessible only from one side and should be _________ if accessible from both sides. The distance between each layer of the shelves shall be _________.
405 mm , 1 m , 455 mm
(GROWTH ROOM) Aisles and passage shall be at least ________.
1 m
(LIGHTING) Outside the laboratory.
100 lux
(LIGHTING) General
200 lux
(LIGHTING) Working table
500 lux
(LIGHTING) Growth room