Pressures Flashcards
(58 cards)
Central Venous Pressure Measurement by Water Manometer
Central venous pressure (CVP) is a luminal pressure measurement taken from the intrathoracic portions of the cranial vena cava. The accepted normal range for CVP is 0 to 10 cm H2O.
Supplies Needed for Central Venous Pressure Measurement
A catheter with the tip properly placed in the intrathoracic vena cava is required. The materials needed include a sterile bag of fluids with an attached fluid administration set, an IV fluid extension set, a three-way stopcock, and a water manometer.
Procedure
The dog or cat should be positioned in lateral or sternal recumbency. A water manometer is placed in the fluid line via a three-way stopcock and the extension set
Central Venous Pressure Measurement by Water Manometer
Central venous pressure (CVP) is a luminal pressure measurement taken from the intrathoracic portions of the cranial vena cava. The accepted normal range for CVP is 0 to 10 cm H2O.
Supplies Needed for Central Venous Pressure Measurement
A catheter with the tip properly placed in the intrathoracic vena cava is required. The materials needed include a sterile bag of fluids with an attached fluid administration set, an IV fluid extension set, a three-way stopcock, and a water manometer.
Procedure
The dog or cat should be positioned in lateral or sternal recumbency. A water manometer is placed in the fluid line via a three-way stopcock and the extension set
The fluid administration set and the extension set are primed with sterile fluids from the fluid bag while the stopcock is in the “Off” position to the water manometer. There must be no air bubbles in the fluid lines. The primed extension set is attached to the patient’s catheter. The stopcock at the bottom of the manometer should rest on the table or cage floor. The administration set is opened, again with the stopcock in the “Off” position to the manometer, to allow fluid to flow into the patient’s catheter, thus ensuring that the catheter is patent. If fluid does not flow freely into the patient’s catheter, a valid CVP measurement will not be obtained. Next, the stopcock is turned so that the “Off” position is now to the patient. The fluid will now fill the manometer, which is allowed to fill about three-quarters full. The stopcock then is turned “Off” to the fluids, allowing a pathway only from the fluid-filled manometer and the patient’s catheter.
The fluid administration set and the extension set are primed with sterile fluids from the fluid bag while the stopcock is in the “Off” position to the water manometer. There must be no air bubbles in the fluid lines. The primed extension set is attached to the patient’s catheter. The stopcock at the bottom of the manometer should rest on the table or cage floor. The administration set is opened, again with the stopcock in the “Off” position to the manometer, to allow fluid to flow into the patient’s catheter, thus ensuring that the catheter is patent. If fluid does not flow freely into the patient’s catheter, a valid CVP measurement will not be obtained. Next, the stopcock is turned so that the “Off” position is now to the patient. The fluid will now fill the manometer, which is allowed to fill about three-quarters full. The stopcock then is turned “Off” to the fluids, allowing a pathway only from the fluid-filled manometer and the patient’s catheter.
The level of fluid in the manometer falls until the hydrostatic pressure of the column of fluid reaches equilibrium with the hydrostatic pressure of the blood at the end of the catheter. Therefore it is essential to know where the catheter tip lies in relation to the manometer fluid column. When the animal is in lateral recumbency, the cranial vena cava lies near the midline, and the manubrium is a good reference point. When the animal is in sternal recumbency, the cranial vena cava is about at the level of the point of the shoulder or the scapulohumeral joint. With the stopcock resting on the tabletop or cage floor, a zero reference point is determined by a horizontal line drawn between the manometer and the appropriate external anatomic landmark (manubrium or scapulohumeral joint). The centimeter mark where the horizontal line intersects the manometer is the zero reference point. A carpenter’s level fastened to a taut string is an excellent guide for the horizontal line from the anatomic landmark to the manometer. The difference between the equilibrium point reading and the zero reference point is the CVP measurement. For example, if the initial reading is 15 cm H2O (where the fluid level stopped falling) and the zero reference point is 10 cm H2O (horizontal line drawn between the external anatomic landmark and the manometer), the CVP is 5 cm H2O. The presence of a well-placed, unobstructed catheter can be verified by the fluid column in the manometer, which will slightly oscillate up and down as the animal’s heart beats or as the animal breathes. Readings are taken between ventilatory excursions. Trends in central venous pressure are more informative than single values. Each time a CVP measurement is obtained, the patient should be in the same recumbent position. The goal is to be consistent in taking the readings and to monitor the trends.
The level of fluid in the manometer falls until the hydrostatic pressure of the column of fluid reaches equilibrium with the hydrostatic pressure of the blood at the end of the catheter. Therefore it is essential to know where the catheter tip lies in relation to the manometer fluid column. When the animal is in lateral recumbency, the cranial vena cava lies near the midline, and the manubrium is a good reference point. When the animal is in sternal recumbency, the cranial vena cava is about at the level of the point of the shoulder or the scapulohumeral joint. With the stopcock resting on the tabletop or cage floor, a zero reference point is determined by a horizontal line drawn between the manometer and the appropriate external anatomic landmark (manubrium or scapulohumeral joint). The centimeter mark where the horizontal line intersects the manometer is the zero reference point. A carpenter’s level fastened to a taut string is an excellent guide for the horizontal line from the anatomic landmark to the manometer. The difference between the equilibrium point reading and the zero reference point is the CVP measurement. For example, if the initial reading is 15 cm H2O (where the fluid level stopped falling) and the zero reference point is 10 cm H2O (horizontal line drawn between the external anatomic landmark and the manometer), the CVP is 5 cm H2O. The presence of a well-placed, unobstructed catheter can be verified by the fluid column in the manometer, which will slightly oscillate up and down as the animal’s heart beats or as the animal breathes. Readings are taken between ventilatory excursions. Trends in central venous pressure are more informative than single values. Each time a CVP measurement is obtained, the patient should be in the same recumbent position. The goal is to be consistent in taking the readings and to monitor the trends.
Evaluation of Hemostasis
Examination of a dog or cat suspected of having a bleeding disorder includes assessment of vascular response and platelet function with formation of the platelet plug (primary hemostasis) and of the coagulation cascade leading to formation of fibrin and stabilization of the plug (secondary hemostasis). Persistent hemorrhage from superficial cutaneous injuries and mucosal surfaces (nasal, gingival, gastrointestinal, genitourinary) and formation of petechiae or ecchymoses are characteristic indicators of abnormalities in primary hemostasis. Primary hemostasis can be evaluated through quantitative measures (platelet estimate from blood smear, platelet count, mean platelet volume) and qualitative measures (von Willebrand factor [vWF] concentration and in vitro platelet function tests, such as optical aggregometry, aperture closure time, and flow cytometry). The buccal mucosal bleeding time (BMBT) is an in vivo, easily performed screening test for evaluation of primary hemostasis in dogs and cats.
Evaluation of Hemostasis
Examination of a dog or cat suspected of having a bleeding disorder includes assessment of vascular response and platelet function with formation of the platelet plug (primary hemostasis) and of the coagulation cascade leading to formation of fibrin and stabilization of the plug (secondary hemostasis). Persistent hemorrhage from superficial cutaneous injuries and mucosal surfaces (nasal, gingival, gastrointestinal, genitourinary) and formation of petechiae or ecchymoses are characteristic indicators of abnormalities in primary hemostasis. Primary hemostasis can be evaluated through quantitative measures (platelet estimate from blood smear, platelet count, mean platelet volume) and qualitative measures (von Willebrand factor [vWF] concentration and in vitro platelet function tests, such as optical aggregometry, aperture closure time, and flow cytometry). The buccal mucosal bleeding time (BMBT) is an in vivo, easily performed screening test for evaluation of primary hemostasis in dogs and cats.
The traditional coagulation pathway has been updated so that it more accurately reflects the in vivo mechanisms of secondary hemostasis (modern coagulation cascade). Abnormalities in secondary hemostasis can cause hematomas, hemarthrosis, and bleeding into the pleural and peritoneal cavities. These abnormalities are identified by coagulation screening tests (prothrombin time, activated partial thromboplastin time, activated coagulation time, fibrinogen), proteins induced by vitamin K absence or antagonism (PIVKA), and individual coagulation factor assays.
The traditional coagulation pathway has been updated so that it more accurately reflects the in vivo mechanisms of secondary hemostasis (modern coagulation cascade). Abnormalities in secondary hemostasis can cause hematomas, hemarthrosis, and bleeding into the pleural and peritoneal cavities. These abnormalities are identified by coagulation screening tests (prothrombin time, activated partial thromboplastin time, activated coagulation time, fibrinogen), proteins induced by vitamin K absence or antagonism (PIVKA), and individual coagulation factor assays.
Published reference ranges for the BMBT are 1.4 to 3.5 minutes in dogs and 1.5 to 2.5 minutes in cats. For clinical purposes in dogs, a BMBT of less than 4 minutes is considered normal.
Published reference ranges for the BMBT are 1.4 to 3.5 minutes in dogs and 1.5 to 2.5 minutes in cats. For clinical purposes in dogs, a BMBT of less than 4 minutes is considered normal.
BMBT measurement is not indicated in animals with thrombocytopenia (less than 70,000/µL), because an abnormal result would be expected. Other causes of a prolonged BMBT are von Willebrand disease (vWD), thrombocytopathies, and vascular disorders. Impaired platelet adhesion and aggregation are present in vWD, a hereditary deficiency of vWF that is reported in various dog breeds but is rare in cats. In a dog with a normal plasma vWF concentration and absence of thrombocytopenia, a prolonged BMBT suggests a defect in platelet function that can be congenital or acquired. Inherited thrombocytopathies (intrinsic platelet function defects) are rare. They include disorders of platelet membranes (e.g., Glanzmann’s thrombasthenia in Otterhounds and Great Pyrenees dogs) and disorders of platelet secretion (e.g., Spitz and Basset Hound thrombopathy and platelet granule storage pool deficiency in American Cocker Spaniels and feline Chédiak-Higashi syndrome). These defects have been associated with a prolonged BMBT. However, not all conditions or agents that cause a platelet dysfunction are reported to result in an abnormal BMBT.
BMBT measurement is not indicated in animals with thrombocytopenia (less than 70,000/µL), because an abnormal result would be expected. Other causes of a prolonged BMBT are von Willebrand disease (vWD), thrombocytopathies, and vascular disorders. Impaired platelet adhesion and aggregation are present in vWD, a hereditary deficiency of vWF that is reported in various dog breeds but is rare in cats. In a dog with a normal plasma vWF concentration and absence of thrombocytopenia, a prolonged BMBT suggests a defect in platelet function that can be congenital or acquired. Inherited thrombocytopathies (intrinsic platelet function defects) are rare. They include disorders of platelet membranes (e.g., Glanzmann’s thrombasthenia in Otterhounds and Great Pyrenees dogs) and disorders of platelet secretion (e.g., Spitz and Basset Hound thrombopathy and platelet granule storage pool deficiency in American Cocker Spaniels and feline Chédiak-Higashi syndrome). These defects have been associated with a prolonged BMBT. However, not all conditions or agents that cause a platelet dysfunction are reported to result in an abnormal BMBT.
Acquired thrombocytopathies are seen in uremic animals, in whom the prolonged BMBT is primarily due to defective platelet adhesion. Platelet aggregation is not consistently altered in uremic animals. Uremic toxins are thought to play a role in the pathogenesis of this condition. A prolonged BMBT compared with pretreatment measurements was noted after intravenous infusion of dextran 70 in dogs. Nonsteroidal antiinflammatory agents interfere with platelet function through inhibition of cyclooxygenase 1 (COX-1), leading to decreased synthesis of thromboxane A2, a platelet-aggregating and vasoconstricting factor. Aspirin was found to prolong the canine BMBT. Carprofen and meloxicam both have COX-1–sparing properties and did not prolong the BMBT in dogs, although platelet aggregation in response to adenosine diphosphate (measured by use of an aggregometer) was decreased after treatment with carprofen, meloxicam, and aspirin in dogs with osteoarthritis. Administration of etodolac, firocoxib, or indomethacin also did not affect the BMBT in healthy dogs. Reports on ketoprofen have been conflicting. Acquired thrombocytopathies associated with dysproteinemias and with disseminated intravascular coagulation are thought to result from platelet coating by paraproteins and fibrin degradation products. Other causes of a prolonged BMBT include infectious diseases (e.g., feline retrovirus-induced thrombocytopathy, canine leishmaniasis, infections with Ehrlichia canis or Anaplasma platys), myeloproliferative diseases, and vasculopathies (e.g., vasculitis and inherited vascular defects). Coagulation factor deficiencies do not produce an abnormal BMBT because functional primary hemostasis leads to the formation of an unstable platelet plug; however, rebleeding can occur.
Acquired thrombocytopathies are seen in uremic animals, in whom the prolonged BMBT is primarily due to defective platelet adhesion. Platelet aggregation is not consistently altered in uremic animals. Uremic toxins are thought to play a role in the pathogenesis of this condition. A prolonged BMBT compared with pretreatment measurements was noted after intravenous infusion of dextran 70 in dogs. Nonsteroidal antiinflammatory agents interfere with platelet function through inhibition of cyclooxygenase 1 (COX-1), leading to decreased synthesis of thromboxane A2, a platelet-aggregating and vasoconstricting factor. Aspirin was found to prolong the canine BMBT. Carprofen and meloxicam both have COX-1–sparing properties and did not prolong the BMBT in dogs, although platelet aggregation in response to adenosine diphosphate (measured by use of an aggregometer) was decreased after treatment with carprofen, meloxicam, and aspirin in dogs with osteoarthritis. Administration of etodolac, firocoxib, or indomethacin also did not affect the BMBT in healthy dogs. Reports on ketoprofen have been conflicting. Acquired thrombocytopathies associated with dysproteinemias and with disseminated intravascular coagulation are thought to result from platelet coating by paraproteins and fibrin degradation products. Other causes of a prolonged BMBT include infectious diseases (e.g., feline retrovirus-induced thrombocytopathy, canine leishmaniasis, infections with Ehrlichia canis or Anaplasma platys), myeloproliferative diseases, and vasculopathies (e.g., vasculitis and inherited vascular defects). Coagulation factor deficiencies do not produce an abnormal BMBT because functional primary hemostasis leads to the formation of an unstable platelet plug; however, rebleeding can occur.
The BMBT is a quick, useful cage-side test for evaluation of primary hemostasis. It requires minimal preparation and few materials, and it is fairly noninvasive, safe, cost-effective, and easy to perform. Results are available immediately, which allows efficient presurgical screening and repeated evaluation to assess response to treatment. The test can be used as a complement to in vitro function tests, which may not fully reflect platelet function in vivo. However, the BMBT procedure is affected by iatrogenic variables that could normalize the result. Reports on sensitivity and specificity for the detection of primary hemostatic disorders vary, and some studies question the sensitivity of the BMBT measurement. A normal result does not eliminate a diagnosis of vWD or a platelet function defect, especially if the degree of dysfunction is mild. No correlation was found between the BMBT and plasma vWF concentration in dogs. The BMBT test is not standardized (e.g., degree of venostasis) and not reproducible. For any two readings in the same dog, the BMBT may vary by up to 2 minutes with one observer or between two observers.
The BMBT is a quick, useful cage-side test for evaluation of primary hemostasis. It requires minimal preparation and few materials, and it is fairly noninvasive, safe, cost-effective, and easy to perform. Results are available immediately, which allows efficient presurgical screening and repeated evaluation to assess response to treatment. The test can be used as a complement to in vitro function tests, which may not fully reflect platelet function in vivo. However, the BMBT procedure is affected by iatrogenic variables that could normalize the result. Reports on sensitivity and specificity for the detection of primary hemostatic disorders vary, and some studies question the sensitivity of the BMBT measurement. A normal result does not eliminate a diagnosis of vWD or a platelet function defect, especially if the degree of dysfunction is mild. No correlation was found between the BMBT and plasma vWF concentration in dogs. The BMBT test is not standardized (e.g., degree of venostasis) and not reproducible. For any two readings in the same dog, the BMBT may vary by up to 2 minutes with one observer or between two observers.
Doppler ultrasonic flow detection through the use of a piezoelectric crystal allows detection of flow in a peripheral artery. Hair is clipped just proximal to the palmar metacarpal pad at the level of the superficial palmar arterial arch for forelimb measurement; over the dorsal pedal artery for hindlimb measurement; or on the ventral aspect of the tail for tail measurement. An occluding cuff (sized as outlined for oscillometric techniques) is placed proximal to the point of flow detection (midradius in the forelimb, proximal to the hock in the hindlimb, or proximal to transducer placement on the tail), and measurements are obtained with the cuff at the level of the heart. Ultrasonic coupling gel is placed on the concave surface of the Doppler transducer, and the transducer is held in position during measurements (Figure 104-4) or fixed in position using adhesive tape.
An audible pulse signal is obtained, and the cuff is inflated with a bulb attached to a pressure gauge. The cuff is inflated to a pressure no less than 40 mm Hg above the audible cutoff point of the signal. The cuff is then slowly deflated, and the pressure at which the Doppler signal is again audible is recorded as the systolic pressure. The cuff is deflated further, and the pressure at which the audible signal abruptly changes in pitch or becomes muffled is recorded as the diastolic pressure.
Doppler ultrasonic flow detection through the use of a piezoelectric crystal allows detection of flow in a peripheral artery. Hair is clipped just proximal to the palmar metacarpal pad at the level of the superficial palmar arterial arch for forelimb measurement; over the dorsal pedal artery for hindlimb measurement; or on the ventral aspect of the tail for tail measurement. An occluding cuff (sized as outlined for oscillometric techniques) is placed proximal to the point of flow detection (midradius in the forelimb, proximal to the hock in the hindlimb, or proximal to transducer placement on the tail), and measurements are obtained with the cuff at the level of the heart. Ultrasonic coupling gel is placed on the concave surface of the Doppler transducer, and the transducer is held in position during measurements (Figure 104-4) or fixed in position using adhesive tape.
An audible pulse signal is obtained, and the cuff is inflated with a bulb attached to a pressure gauge. The cuff is inflated to a pressure no less than 40 mm Hg above the audible cutoff point of the signal. The cuff is then slowly deflated, and the pressure at which the Doppler signal is again audible is recorded as the systolic pressure. The cuff is deflated further, and the pressure at which the audible signal abruptly changes in pitch or becomes muffled is recorded as the diastolic pressure.
The Doppler flow detection system of BP measurement is considered the most accurate and repeatable of studied noninvasive BP measurement techniques in conscious cats. It is frequently used in dogs but may provide spurious high readings in some individuals. When abnormal readings are obtained by this method from dogs, care should be taken to make sure the abnormal readings are repeatable over multiple measurement periods. The advantages of this technique include flexibility with regard to motion, low pressure, small vessels, or the presence of arrhythmias, as well as speed of measurement. The rapidity of the measurement techniques allows for prompt assessment of changing BP, but meticulous attention must be paid to obtaining strong, audible signals in order to obtain the most accurate BP readings. This technique is also the most operator dependent of the techniques discussed. Accurate identification of diastolic pressures improves with operator practice, and BP measurement using Doppler ultrasonic flow detection techniques is most accurate if a few well-trained individuals in a practice are responsible for this diagnostic test and perform the test frequently.
The Doppler flow detection system of BP measurement is considered the most accurate and repeatable of studied noninvasive BP measurement techniques in conscious cats. It is frequently used in dogs but may provide spurious high readings in some individuals. When abnormal readings are obtained by this method from dogs, care should be taken to make sure the abnormal readings are repeatable over multiple measurement periods. The advantages of this technique include flexibility with regard to motion, low pressure, small vessels, or the presence of arrhythmias, as well as speed of measurement. The rapidity of the measurement techniques allows for prompt assessment of changing BP, but meticulous attention must be paid to obtaining strong, audible signals in order to obtain the most accurate BP readings. This technique is also the most operator dependent of the techniques discussed. Accurate identification of diastolic pressures improves with operator practice, and BP measurement using Doppler ultrasonic flow detection techniques is most accurate if a few well-trained individuals in a practice are responsible for this diagnostic test and perform the test frequently.
Arterial Puncture or Arterial Cannulation (“Invasive” or “Direct” Technique)
Direct BP measurement involves advancing a needle attached to a pressure transducer directly into an artery to measure BP. This procedure may be performed acutely to obtain instantaneous BP readings, or BP can be measured over time through the use of an indwelling arterial catheter instead of a needle. Invasive BP measurements are typically used during anesthetic procedures, in critical care patients when ongoing BP information is desired, or to document or exclude hypertension acutely as a clinical diagnosis in dogs. Acute arterial puncture is seldom used for clinical diagnosis of hypertension in cats.
Direct BP measurement is usually performed by means of puncture of the femoral artery in dogs. Use of local anesthesia is strongly recommended for this procedure and when used, the procedure is well tolerated by most dogs. The patient is gently restrained in lateral recumbency. Approximately 5 minutes prior to arterial puncture, 1 to 2 mL of 2% lidocaine hydrochloride is injected subcutaneously over the area in which the femoral pulse is palpated. A 22-gauge, 1-inch needle is attached to a transducer and flushed with heparinized saline, ensuring that no bubbles remain in the transducer. The transducer is zeroed at the level of the sternum in the laterally recumbent dog. The femoral pulse is palpated in the femoral triangle, and the needle is carefully advanced into the femoral artery (Figure 104-1) until a satisfactory pressure waveform is recorded on the monitor screen.
Arterial Puncture or Arterial Cannulation (“Invasive” or “Direct” Technique)
Direct BP measurement involves advancing a needle attached to a pressure transducer directly into an artery to measure BP. This procedure may be performed acutely to obtain instantaneous BP readings, or BP can be measured over time through the use of an indwelling arterial catheter instead of a needle. Invasive BP measurements are typically used during anesthetic procedures, in critical care patients when ongoing BP information is desired, or to document or exclude hypertension acutely as a clinical diagnosis in dogs. Acute arterial puncture is seldom used for clinical diagnosis of hypertension in cats.
Direct BP measurement is usually performed by means of puncture of the femoral artery in dogs. Use of local anesthesia is strongly recommended for this procedure and when used, the procedure is well tolerated by most dogs. The patient is gently restrained in lateral recumbency. Approximately 5 minutes prior to arterial puncture, 1 to 2 mL of 2% lidocaine hydrochloride is injected subcutaneously over the area in which the femoral pulse is palpated. A 22-gauge, 1-inch needle is attached to a transducer and flushed with heparinized saline, ensuring that no bubbles remain in the transducer. The transducer is zeroed at the level of the sternum in the laterally recumbent dog. The femoral pulse is palpated in the femoral triangle, and the needle is carefully advanced into the femoral artery (Figure 104-1) until a satisfactory pressure waveform is recorded on the monitor screen.
A sample of the tracing is recorded, and the needle is withdrawn. Firm pressure is applied to the area of arterial puncture for a minimum of 5 minutes after measurement. The patient should be monitored closely for at least 1 hour after the procedure for any complications related to hematoma formation. Systolic, diastolic, and mean BP values from five consecutive cardiac cycles during normal sinus rhythm are averaged to obtain a representative value for the patient. When use of an arterial catheter is preferred, the catheter is usually inserted into the dorsal pedal artery. A local anesthetic may be used as described previously.
A sample of the tracing is recorded, and the needle is withdrawn. Firm pressure is applied to the area of arterial puncture for a minimum of 5 minutes after measurement. The patient should be monitored closely for at least 1 hour after the procedure for any complications related to hematoma formation. Systolic, diastolic, and mean BP values from five consecutive cardiac cycles during normal sinus rhythm are averaged to obtain a representative value for the patient. When use of an arterial catheter is preferred, the catheter is usually inserted into the dorsal pedal artery. A local anesthetic may be used as described previously.
Oscillometric Technique
Oscillometric BP measurement involves the use of an automated detection system and a cuff that is wrapped around a limb or tail over an artery. The cuff is inflated automatically to a pressure that causes occlusion of the artery and then slowly deflated. The machine detects oscillations in the vessel wall as the occlusion is eased, and the pressure at which oscillations are maximal is recorded as the mean arterial pressure. Systolic and diastolic pressures are then calculated by the monitor using algorithms specific to the technique. This technique uses data from many cardiac cycles to render a single reading and is therefore unsuitable for use in animals with rapidly changing BP.
Oscillometric BP measurement methods are more accurate in dogs than in cats. This measurement method may return inaccurate results if the patient is not motionless (e.g., trembling), has a weak or irregular pulse, or has a small artery (i.e., most cats). Use of a cuff of appropriate size is extremely important to ensure accurate measurements. Cuff width should be approximately 40% of the circumference of the limb or tail in dogs and approximately 30% of the appendage circumference in cats. When used on the tail, the cuff is wrapped snugly high on the tail head with the dog in sternal or lateral recumbency. Although tail cuffs can be used in standing animals, animal movement often interferes with accurate measurement. Limb cuffs are wrapped around the forelimb distal to the elbow or around the midmetatarsus at the level of the superficial plantar arterial arch.
To maximize accuracy, the cuff should be at the level of the heart during readings; therefore lateral or sternal recumbency is preferred, and use in standing patients is not recommended. In cats, tail cuffs return more repeatable measurements than limb cuffs. Typically, the cat rests in sternal recumbency during readings
Oscillometric Technique
Oscillometric BP measurement involves the use of an automated detection system and a cuff that is wrapped around a limb or tail over an artery. The cuff is inflated automatically to a pressure that causes occlusion of the artery and then slowly deflated. The machine detects oscillations in the vessel wall as the occlusion is eased, and the pressure at which oscillations are maximal is recorded as the mean arterial pressure. Systolic and diastolic pressures are then calculated by the monitor using algorithms specific to the technique. This technique uses data from many cardiac cycles to render a single reading and is therefore unsuitable for use in animals with rapidly changing BP.
Oscillometric BP measurement methods are more accurate in dogs than in cats. This measurement method may return inaccurate results if the patient is not motionless (e.g., trembling), has a weak or irregular pulse, or has a small artery (i.e., most cats). Use of a cuff of appropriate size is extremely important to ensure accurate measurements. Cuff width should be approximately 40% of the circumference of the limb or tail in dogs and approximately 30% of the appendage circumference in cats. When used on the tail, the cuff is wrapped snugly high on the tail head with the dog in sternal or lateral recumbency. Although tail cuffs can be used in standing animals, animal movement often interferes with accurate measurement. Limb cuffs are wrapped around the forelimb distal to the elbow or around the midmetatarsus at the level of the superficial plantar arterial arch.
To maximize accuracy, the cuff should be at the level of the heart during readings; therefore lateral or sternal recumbency is preferred, and use in standing patients is not recommended. In cats, tail cuffs return more repeatable measurements than limb cuffs. Typically, the cat rests in sternal recumbency during readings
In all cases, best results are obtained when the patient is minimally restrained and soothed during the procedure. A short acclimation period prior to measurement is recommended to allow the cat to become more calm. A series of at least five readings are obtained at approximately 1-minute intervals. Any readings that are clearly erroneous are discarded. The multiple readings are then averaged to obtain a representative result.
Oscillometric techniques are valuable in anesthetized dogs and cats to monitor trends in BP. Because the animal is immobilized and BP shows less variability over time, repeatable and accurate readings can be obtained over time in both dogs and cats.
In all cases, best results are obtained when the patient is minimally restrained and soothed during the procedure. A short acclimation period prior to measurement is recommended to allow the cat to become more calm. A series of at least five readings are obtained at approximately 1-minute intervals. Any readings that are clearly erroneous are discarded. The multiple readings are then averaged to obtain a representative result.
Oscillometric techniques are valuable in anesthetized dogs and cats to monitor trends in BP. Because the animal is immobilized and BP shows less variability over time, repeatable and accurate readings can be obtained over time in both dogs and cats.
Effect of body position on indirect measurement of systolic arterial blood pressure in dogs
A diagnostic test evaluation to determine whether a difference existed in Doppler ultrasonographic measurements of systolic arterial blood pressure (SAP) in sitting versus laterally recumbent dogs and to determine the degree of variability in measurements made in each position.
In a crossover design, SAP was measured in 51 healthy or sick adult dogs, without recent sedation or anesthesia and with an SAP ≤ 300 mm Hg via Doppler ultrasonography when dogs were sitting (on hind limbs with nonmeasured forelimb bearing weight) and laterally recumbent, with the cuff position at the level of the right atrium for both positions. Seven measurements were obtained per position for each dog.
Mean ± SD SAP was significantly higher in the sitting (172.1 ± 33.3 mm Hg) versus recumbent (147.0 ± 24.6 mm Hg) position, and this difference was evident for 44 of 51 (86%) dogs. The mean difference in measured SAP between the 2 positions was 25.1 ± 28.5 mm Hg. Blood pressure measurements had a significantly higher repeatability in the recumbent position than in the sitting position.
Blood pressure measurements in dogs were significantly affected by body position, and they were higher for most dogs when sitting rather than laterally recumbent. Blood pressure measurements in the laterally recumbent body position were less variable than in the sitting position.
Effect of body position on indirect measurement of systolic arterial blood pressure in dogs
A diagnostic test evaluation to determine whether a difference existed in Doppler ultrasonographic measurements of systolic arterial blood pressure (SAP) in sitting versus laterally recumbent dogs and to determine the degree of variability in measurements made in each position.
In a crossover design, SAP was measured in 51 healthy or sick adult dogs, without recent sedation or anesthesia and with an SAP ≤ 300 mm Hg via Doppler ultrasonography when dogs were sitting (on hind limbs with nonmeasured forelimb bearing weight) and laterally recumbent, with the cuff position at the level of the right atrium for both positions. Seven measurements were obtained per position for each dog.
Mean ± SD SAP was significantly higher in the sitting (172.1 ± 33.3 mm Hg) versus recumbent (147.0 ± 24.6 mm Hg) position, and this difference was evident for 44 of 51 (86%) dogs. The mean difference in measured SAP between the 2 positions was 25.1 ± 28.5 mm Hg. Blood pressure measurements had a significantly higher repeatability in the recumbent position than in the sitting position.
Blood pressure measurements in dogs were significantly affected by body position, and they were higher for most dogs when sitting rather than laterally recumbent. Blood pressure measurements in the laterally recumbent body position were less variable than in the sitting position.
A pulse oximeter is a device designed to measure the percent saturation of arterial hemoglobin with oxygen. The value calculated is usually referred to as SpO2 to differentiate it from a value obtained by direct measurement from arterial blood (SaO2). To understand the likelihood of this device producing an accurate result it is important to understand how it works.
A pulse oximeter is a device designed to measure the percent saturation of arterial hemoglobin with oxygen. The value calculated is usually referred to as SpO2 to differentiate it from a value obtained by direct measurement from arterial blood (SaO2). To understand the likelihood of this device producing an accurate result it is important to understand how it works.
Principles of pulse oximetry:
The basis for this technology is that oxygenated hemoglobin and deoxygenated hemoglobin have different light absorption characteristics and hence different colors. Light absorption is usually measured as an extinction coefficient (ε), which was originally defined as the thickness of the material that was required to decrease the intensity of light passing though it to one tenth of the incident light.1 This is depicted in Figure 105-1 showing ε for hemoglobin (Hb), oxyhemoglobin (HbO2), methemoglobin (MetHb), and carboxyhemoglobin (HbCO) over a spectrum of light wavelengths from orange (600 nm) through to infrared (>700 nm).
Principles of pulse oximetry:
The basis for this technology is that oxygenated hemoglobin and deoxygenated hemoglobin have different light absorption characteristics and hence different colors. Light absorption is usually measured as an extinction coefficient (ε), which was originally defined as the thickness of the material that was required to decrease the intensity of light passing though it to one tenth of the incident light.1 This is depicted in Figure 105-1 showing ε for hemoglobin (Hb), oxyhemoglobin (HbO2), methemoglobin (MetHb), and carboxyhemoglobin (HbCO) over a spectrum of light wavelengths from orange (600 nm) through to infrared (>700 nm).
Pulse oximeters are a standard of care in human anesthesia and critical care. There is some controversy over the true benefit of this requirement but it would be difficult to get an estimate of its impact since it would now be almost impossible to eliminate access to this technology.13-16 Most veterinary practices have at least one pulse oximeter, but it appears that understanding the information provided by the device is deficient.17 It is important to recognize that pulse oximeters do not measure the amount of Hb present or the quantity of blood flow. It is quite possible for saturation to be normal but oxygen delivery inadequate if Hb concentration and/or cardiac output are low.
Pulse oximeters are a standard of care in human anesthesia and critical care. There is some controversy over the true benefit of this requirement but it would be difficult to get an estimate of its impact since it would now be almost impossible to eliminate access to this technology.13-16 Most veterinary practices have at least one pulse oximeter, but it appears that understanding the information provided by the device is deficient.17 It is important to recognize that pulse oximeters do not measure the amount of Hb present or the quantity of blood flow. It is quite possible for saturation to be normal but oxygen delivery inadequate if Hb concentration and/or cardiac output are low.
Pulse oximetry in anesthetized animals would appear to be good for monitoring desaturation. However, when one examines the oxyhemoglobin dissociation curve (Figure 105-4), it is clear that Hb becomes almost fully saturated at a PaO2 of about 100 mm Hg, which is achievable when breathing room air.
Pulse oximetry in anesthetized animals would appear to be good for monitoring desaturation. However, when one examines the oxyhemoglobin dissociation curve (Figure 105-4), it is clear that Hb becomes almost fully saturated at a PaO2 of about 100 mm Hg, which is achievable when breathing room air.
When an animal is breathing a high concentration of oxygen and has a PaO2 of 500 mm Hg, the pulse oximeter should read 99% to 100%, and if something significant happens to the animal (e.g., pneumothorax, pulmonary edema, atelectasis) that decreases the PaO2 to ~250 mm Hg, the pulse oximeter will likely continue to read 99% to 100%, and it will not get to an SpO2 of 95% until PaO2 gets to about 90 mm Hg.20 This makes the tool a crude monitor for healthy animals breathing >90% O2, and the problem is compounded by the lack of reliability of the technology. It is common to see pulse oximeters giving SpO2 values of
When an animal is breathing a high concentration of oxygen and has a PaO2 of 500 mm Hg, the pulse oximeter should read 99% to 100%, and if something significant happens to the animal (e.g., pneumothorax, pulmonary edema, atelectasis) that decreases the PaO2 to ~250 mm Hg, the pulse oximeter will likely continue to read 99% to 100%, and it will not get to an SpO2 of 95% until PaO2 gets to about 90 mm Hg.20 This makes the tool a crude monitor for healthy animals breathing >90% O2, and the problem is compounded by the lack of reliability of the technology. It is common to see pulse oximeters giving SpO2 values of
Measurement of Blood Pressure
Since the pulse oximeter can display a pulse signal, it can be placed on a distal extremity and a cuff placed above this site to occlude blood flow. The cuff can be inflated to a point above systolic arterial blood pressure and then deflated while watching the plethysmographic trace on the pulse oximeter. Systolic pressure should be the pressure in the cuff where the trace returns. In one study, in cats, the pressure measured by this method had a greater correlation to mean pressure than to systolic pressure.21 In dogs using the tongue as the measurement site the technique was neither accurate nor precise.22
Measurement of Blood Pressure
Since the pulse oximeter can display a pulse signal, it can be placed on a distal extremity and a cuff placed above this site to occlude blood flow. The cuff can be inflated to a point above systolic arterial blood pressure and then deflated while watching the plethysmographic trace on the pulse oximeter. Systolic pressure should be the pressure in the cuff where the trace returns. In one study, in cats, the pressure measured by this method had a greater correlation to mean pressure than to systolic pressure.21 In dogs using the tongue as the measurement site the technique was neither accurate nor precise.22
Systemic hypertension (HT) refers to the persistent elevation of systemic blood pressure (BP). Systemic hypertension is an increasingly recognized source of morbidity and, in some cases, mortality in human and veterinary patients.1-7 Overt and sometimes devastating damage caused by HT is typically noted in the eyes,5,8 central nervous system,9,10 heart,7,11-13 and kidneys.2,6,14 Injury related to HT in these organ systems is often collectively termed target organ damage (TOD). TOD is often clinically obvious, especially in the ocular or nervous systems. In some cases, however, TOD can be insidious, resulting in undetected progressive deterioration of damaged organs (e.g., accelerated deterioration of renal function6,9).
Systemic hypertension (HT) refers to the persistent elevation of systemic blood pressure (BP). Systemic hypertension is an increasingly recognized source of morbidity and, in some cases, mortality in human and veterinary patients.1-7 Overt and sometimes devastating damage caused by HT is typically noted in the eyes,5,8 central nervous system,9,10 heart,7,11-13 and kidneys.2,6,14 Injury related to HT in these organ systems is often collectively termed target organ damage (TOD). TOD is often clinically obvious, especially in the ocular or nervous systems. In some cases, however, TOD can be insidious, resulting in undetected progressive deterioration of damaged organs (e.g., accelerated deterioration of renal function6,9).
Systemic HT is typically subclassified as primary (“essential”) HT or secondary HT. Some veterinarians have attempted to distinguish essential, or primary, HT from idiopathic HT (i.e., systemic HT in the absence of overt, clinically apparent causal disease). Use of the term idiopathic acknowledges that there may be a causal disease (e.g., renal disease) that is responsible for HT but that the causal disease is in a preclinical phase.15 In cases where the underlying disease is rare (e.g., pheochromocytoma), discovery of the cause of the HT is dependent on thorough diagnostic testing. In most dogs and cats, HT is a complication of a systemic disease rather than being primary.
Systemic HT is typically subclassified as primary (“essential”) HT or secondary HT. Some veterinarians have attempted to distinguish essential, or primary, HT from idiopathic HT (i.e., systemic HT in the absence of overt, clinically apparent causal disease). Use of the term idiopathic acknowledges that there may be a causal disease (e.g., renal disease) that is responsible for HT but that the causal disease is in a preclinical phase.15 In cases where the underlying disease is rare (e.g., pheochromocytoma), discovery of the cause of the HT is dependent on thorough diagnostic testing. In most dogs and cats, HT is a complication of a systemic disease rather than being primary.