Surveys Flashcards
(129 cards)
Protects XRT from rough handling
Protective Tube Housing
Lead-lined metal to protect the patient and staff from off-focus radiation
Protective Tube Housing
Must be constructed so that the leakage radiation measured at a distance of 1m from the x-ray source does not exceed 1 mGya/hr when the tube is operated at its highest voltage at the highest.
Protective Tube Housing
Off-focus radn
Leakage radn
Accepted amung of leakage radn from protective tube housing (Give 2 values)
<100mR/hr or <1mGya/hr at 1m distance
It is where the technical exposure factors such
as milliamperes (mA) and
peak kilovoltage (kVp) are selected and visually displayed.
Control Console
Control Console is located (1) that has a/n (2) that permits observation of the patient during any procedure.
- behind a suitable protective barrier
- radiation-absorbent window
Where most of radiographic procedures are performed
Radiographic table
Must be strong and adequately support the pt
Radiographic table
Why should the radiographic table be radiolucent as possible?
Should be as radiolucent as possible so that it will absorb only a minimal amount of radn, thereby reducing the pt’s radn dose
Composition of Radiographic Table
Carbon Fiber
An equipment that measure the distance from the anode focal spot to the IR to ensure that the correct source-to-image receptor distance (SID) is maintained
SID Indicator
A tape erasure attached to the collimator or tube housing or lasers
SID Indicator
SID Indicator Distance and centering indicators must be accurate to within (1) and (2) of the SID respectively
- 2%
- 1%
The SID Indicator measures the distance from (1) to the (2) to ensure correct (3)
- anode focal spot
- IR
- SID
Devices that confine the useful, or primary beam before it enters the area of clinical interest
XR Beam Limiting Devices
XR Beam Limiting Devices reduces the amount of (1) in the tissue and prevents (2) to tissues
- scattered radiation
- unnecessary exposure
XR Beam Limiting Devices
Collimator
Aperture Diaphragm
Cones and Cylinders
Collimator aka
Light-Localizing Variable-Aperture Rectangular Collimator
Light-Localizing Variable-Aperture Rectangular Collimator Consists of these
- Two sets of adjustable lead shutters
- Light source
- Mirror
Differentiate the 2 sets of adjustable lead shutters
- First set/ upper shutters: located as close as possible to the tube window to reduce amount of off-focus radiation
- Second set/ lower shutters: below the level of light source & mirror; further confine beam to area of clinical interest
Describe the light source
Illuminates XR field and permits it to be centered over area of clinical interest
Describe the mirror
Deflects light beam towards pt to be radiographed
Automatically adjusts activated collimators so the radiation field matches the size of the IR
Positive Beam Limitation (PBL)