Match the medical gases to the corresponding color (USA):
Oxygen
Carbon dioxide
Nitrous oxide
Air
Oxygen
Green
Carbon dioxide
Gray
Nitrous oxide
Blue
Air
Yellow
The High-pressure circuit in a High/Intermediate/Low system may include ?
High pressure regulator
tank Bourdon gauge
Hanger Yoke
The purpose of the high pressure regulator is?
To reduce high pressure from a tank to a working pressure
If contamination or a construction crossover of piped medical gases in the operating room is suspected, which of the following actions should be taken?
The machine’s back up cylinder valve should be opened and the anesthesia machine should bbe disconnected from the hospital’s pipeline supply
Tanks on the anesthesia machine should remain closed when utilizing pipeline gases for what reasons?
So the low oxygen pressure alarm will sound if pipeline supply fails, prior to any tank usage
So the anesthetist maintains a reserve supply of oxygen at hand if pipeline supply fails
True or False:
A scavenging system with suction applied is classified as active scavenging.
True
Intermediate pressures in the pipeline system are typically around __________psig, while intermediate pressures from the tank system are typically around ____________psig.
50, 45