A sevoflurane vaporizer will deliver an accurate concentration of an unknown volatile anesthetic if the latter shares which property with sevoflurane?
Vapor Pressure
True or False:
Any concentration calibrated vaporizer is designed to be utilized for every volatile anesthetic except Desflurane.
False
Choose the BEST MAC estimates for the following gases:
ISO 1 SEVO 2 DES 6 N2O 100
If Halothane is placed in an Enflurane vaporizer and Sevoflurane is placed in an Isoflurane vaporizer; Halothane delivered will be __________, and Sevoflurane delivered will be __________.
Higher, Lower
If a carrier gas flow through a vaporizer is increased, gas delivered to the patient _________, and temperature of the liquid __________.
increases,decreases
If a vaporizer is tilted > 45 degrees:
agent in vaporization chamber will likely decrease, agent in bypass chamber will likely increase
If flow is _______ through a vaporizer, thermocompensation or supplied heat needs to _______.
increased, increase
When using a 3 vaporizer manifold type of machine, if one removes the ________ vaporizer the interlock safety system will not work.
middle vaporizer
Which Vaporizer is NOT a variable bypass vaporizer?
Tec 6 Desflurane vaporizer
Which vaporizer does NOT use an internal thermocompensation mechanism? (Heat is provided through an electrical heater)
Tec 6