Saia Unit 3 Flashcards
(143 cards)
the 3 cardinal rules for xray:
time, distance, shielding
inverse square law:
I1/I2=D2^2/D1^2
primary protective barrier:
direct impact, 1/16” lead or equivelent; 7 feet high
secondary protective barrier:
scatter/leakage, 1/32” lead equivelent
xray must scatterat least:
2 times before hitting the control booth
protective tube housing enclosed by lead lined metal covering:
1/16 lead, reduces leakage
leakage from the tube housing should exceed:
100mR/hr at 1 meter
time of occupancy factor:
amount of time an area is ossupied and by whom
workload:
radiation activity level (max kvp and max mas)
Use:
the amount of time xray is “on” during the day
Lead underwear keep radiation and are used for:
brachytherapy of the prostate
mobile unit switch
should allow tech to obtain at least a distance of 6’ from patient
fluoroscopy drapes and bucky slot cover should have a minimum:
.25mm pb/eq
fluoroscopy deadman switch:
only emits radiation when constant pressure is applied
c-arm source to tabletop must not be less than:
12”
source to tabletop for fixed fluoro unit must not be less than:
15”
Total filtration of fluoroscopy equipment must be at least:
2.5mm al/eq
cumulitive timing device in fluoro creates an audible sound after:
5 minutes
lead used in fluoroscopy must be at least:
.5mm pb/eq
Fluoroscopy xray intensity at tabletop must not exceed:
10R/min
Fluoroscopy uses a high KV and low MA to
minimize dose
Fluoroscopy MA must not exceed:
5MA
Air exposure is measured in:
Traditional - Roentgen (R) ; SI - Air Kerma - grays in air (gyA)
Absorbed dose is measured in:
Traditional - RAD; SI - grays in tissue (gyT)