**Saia Unit 4 Flashcards

(231 cards)

1
Q

Energy deposited into the image receptor

A

Density

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
2
Q

The controlling factors of density are

A

mA
Time
mAs

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
3
Q

mAs

A

The number of X-rays in polyenergetic primary beam

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
4
Q

mAs has the primary control of

A

Image density

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
5
Q

Density is placed in one of three categories

A

Acceptable
Underexposed (too light)
Overexposed (too dark)

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
6
Q

30% rule

A

The human eye requires a change in density of at least 30% before it can be visualized

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
7
Q

Fuji exposure indicator

A

S number
Range 180-220
Indirect relationship

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
8
Q

Kodak exposure number

A

EI number
Range 1800-2200
Direct relationship

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
9
Q

Agfa exposure indicator

A

LGM number
Range - 1.9-2.5
Direct relationship

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
10
Q

A perfect exposure indicator value is

A

1mR of exposure to imaging plate/panel

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
11
Q

Increasing the KVP will increase the quality or energy energy of the beam giving the beam more

A

Penetration ability

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
12
Q

As KVP is increased, density…

A

Will increase

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
13
Q

Energy deposited into the image receptor is

A

Density

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
14
Q

15% rule

A

Used to alter KVP settings to change density and contrast
A 15% increase in KVP = a 50% decrease in mAs
A 15% decrease in KVP = a doubling in mAs

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
15
Q

As KVP goes up, image contrast

A

Goes down

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
16
Q

As KVP goes down, image contrast

A

Goes up

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
17
Q

If SID is increased with no other change, density is

A

Decreased

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
18
Q

If SID is decreased with no other change density is

A

Increased

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
19
Q

SID and density have an ? relationship

A

Inverse

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
20
Q

Density

A

Overall blackening of a film/image

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
21
Q

Inverse square law

A

The intensity of the beam is inversely proportional to the square of the distance.

I1 (D2)2/I2 (D1)2

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
22
Q

Density maintenance law

A

Formula that will compensate for changes in SID to maintain density

mAs1 (D1)2/mAs2 (D2)2

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
23
Q

If you increase the OID less scatter reaches the film resulting in a decrease in

A

Density

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
24
Q

Less OID will result in a greater density due to

A

More radiation reaching the film

How well did you know this?
1
Not at all
2
3
4
5
Perfectly
25
OID and density have a ? relationship
Inverse
26
Imaging system speed
As speed increases the efficiency of placing energy into the IR increases, and density increases also.
27
Grids are used to
Absorb scatter to prevent it from reaching the film.
28
Formula to maintain density with grids
Mas1 GCF1 / mAs 2 GCF2
29
If collimation is increased (smaller light field) then
Less radiation leaves the tube Less scatter produced Less energy reaches IR LESS DENSITY IS THE RESULT
30
Radiolucent tissue
Tissue that let's xray energy through
31
Radiopaque
Tissue that will absorb ray energy
32
Disease
Physical and chemical changes in tissue
33
Anode heel effect
Variation in the beams intensity across its longitudinal axis, More X-rays on cathode side than anode side
34
As filtration is increased,
More photons are absorbed resulting in fewer X-rays and less xray energy in the primary beam
35
Contrast
The difference in adjacent shades across the radiographic image
36
Contrasts primary function
To make recorded detail visible
37
Contrast is primarily controlled by
KVP
38
Energy deposited into the IR is
Density The more energy values = more shades of density
39
High contrast
Few shades of gray Increased contrast Lower KVP "Short scale" contrast
40
Low contrast
Many shades of gray Decreased contrast High KVP "Long scale" contrast
41
Differential absorption
Differing materials absorb xray at differing degrees
42
As KVP decreases patient dose ?
Increases (weaker xray, more absorbed in patient)
43
As KVP increases patient dose ?
Decreases | Stronger xray beam, more passes through patient
44
Grid ratio
Height of lead strips to distance between
45
Grid frequency
Number of lead strips per unit distance
46
Increasing beam restriction - smaller field
Less tissue irradiated Less scatter produced - "short scale", higher contrast
47
An increase in OID results in
Less scatter reaching the film - higher contrast
48
As beam filtration increases
Shorter scale | Higher contrast
49
Positive contrast media
(Iodine/barium) - temporarily make anatomy more dense than normal
50
Negative contrast media
Oxygen, carbon dioxide, room air | Allow more X-rays to penetrate due to the low atomic number resulting in an increase in density
51
Barium atomic number
56
52
Iodine atomic number
53
53
High contrast
Big difference in shades, short scale, low KVP, very black and white Think of the top of shade triangle (small peak)
54
Low contrast
Little difference between shades, long scale, high KVP, many grays Think of the bottom of shade triangle (large base)
55
Photographic factors
Density | Contrast
56
Density
Overall image blackness
57
Contrast
Visibility of adjacent structures
58
Geometric factors
Detail | Distortion
59
Detail
Visibility of fine anatomical structures
60
Distortion
Misrepresentation of an objects size/shape
61
Recorded detail aka
Resolution, sharpness, definition, detail Measured in lp/mm
62
Line pair
A line and it's adjacent space, as recorded in an image of the tool (resolution tool)
63
Human visual acuity is
5 lp/mm
64
Unsharpness
Degree of loss of detail of the anatomical structure by the imaging system
65
Penumbra
Geometric unsharpness around the periphery of the structure of the image
66
As focal spot size increases
Penumbra increases
67
Effective focal spot
focal spot projected toward the patient | Usually .5-2.0mm
68
Line focus principal
The effective focal spot is smaller than the true/actual focal spot
69
True/actual focal spot
Area where electron stream strikes the anode target
70
Small focal spot
Maximum image detail
71
Umbra
Objects true size
72
Why not always use a small focal spot
Small focal spot uses small filament which are mA limited, large focal sometimes needed
73
Increased SID will result in
Less penumbra
74
Increased OID will result in
More penumbra
75
Macro radiography
Magnification radiography Intentionally magnifying to exaggerate small structures Most common in mammography
76
IR speed
Faster speed is used to save patient dose, however resolution not good
77
Quantum mottle
Not enough X-rays in the primary beam to adequately form image
78
Distortion
Misrepresentation of an objects true size and shape
79
Size distortion is always
Magnification
80
Magnification
The misrepresentation of the actual size of an object
81
As SID increases, magnification?
Decreases
82
As OID increases, magnification ?
Increases
83
Magnification factor
SID/SOD (SID divided by SOD)
84
SID/SOD
Image width/object width = SID/SOD
85
Shape distortion
Misrepresentation of an anatomical part due to misalignment of source, IR, and/or part Results in elongation or foreshortening
86
When the xray beam or IR is misaligned =
Elongation
87
When the body part is misaligned=
Foreshortening
88
Three points to perfect alignment
CR perpendicular to body part (most important) CR perpendicular to IR Body part parallel to IR
89
Magnification formula
Mag=SID/SOD
90
Percent magnification =
Image size - object size / object size X 100 = % magnification
91
Intensifying screens
Used to convert xray energy into visible light energy. - 90-99% of film exposure is due to visible light - 1-10% of film exposure is due to direct xray
92
Xray film is much more sensitive to
Visible light energy than it is to xray energy
93
Spectral emission
The color of visible light emitted by a phosphor crystal when stimulated
94
Intensifying screen construction:
Base layer Reflective layer Phosphor layer (phosphor crystals) Protective layer
95
Function of radiographic film
To record and store a permanent anatomical image
96
Radiographic film construction
``` Protective layer Emulsion layer (silver halide crystals) Adhesive layer Base layer Adhesive layer Emulsion layer (silver halide crystals) Protective layer ```
97
Density
Energy deposited into an image receptor
98
Emulsion layer
Silver halide crystals - gelatin whose function is to absorb and then release the processing chemistry and water
99
Colloid
A material that can absorb and release water without itself being dissolved in the process. Like a sponge.
100
Function of the silver halide crystals
Respond when stimulated by xray or light energy
101
Latent image
Formed in the films emulsion as a shadow representation of anatomy. Due to chemical changes in AgBr crystals. Invisible to human eye.
102
Manifest
Visible
103
Latent
Hidden
104
Manifest image
Formed in films emulsion due to chemical changes in AgBr crystals, made visible to the human eye by processor.
105
Gurney Mott theory
Photon breaks ionic bond releasing shared electrons which drift to and are captured by speck. Electrons attaching to the development center. Silver ion attracted to the development center. Sensitivity speck is now neutral and able to attract an electron. More energy absorbed = more debris = bigger sensitivity speck
106
Grid conversion
No Grid = 1 5: 1 = 2 6: 1 = 3 8: 1 = 4 12: 1 = 5 16: 1 = 6
107
Filtration
Minimum of 2.5mm Al/Eq
108
Purpose of filtration
Remove weaker X-rays from primary beam
109
Half value layer
Amount of filtration required to lower xray intensity to half of its original volume
110
Collimator illuminance
Brightness of collimator bulb and field | - at least 15 foot candles (160 lux) at 40"
111
Xray to light field
Collimator to xray beam alignment | - +/-2% semiannual
112
Nine penny test
Test done for xray to light field (diameter of penny is .8 inches)
113
Positive beam limitation
Auto collimation of light field to IR size, can be smaller - never larger! - +/-2% semiannual
114
Focal spot size is tested by
Pinhole camera, star pattern or slit camera | - +/-50% annual
115
KVP accuracy is measured with
Voltmeter and radiation meter | - +/-10% annual
116
Exposure timer accuracy
+/-5% greater than 10ms | +/- 20% less than 10ms
117
Exposure linearity
Using the same mAs but different combinations of mA and S | i.e. 100mA @1/2S or 200mA @1/4S
118
AEC backup timer is set to self terminate if
If 600 mAs is reached
119
If AEC backup timer is manually set, it should be set to
150% of expected mAs
120
AEC optical density versus change
+/-.3 OD
121
Film illuminator standard is
15W daylight bulb
122
Fluoro exposure rate ESE shall not exceed
10R/min
123
Interventional exposure rate ESE shall not exceed
20R/min
124
Repeat analysis goal
To minimize patient exposure
125
Wire mesh test is used to evaluate
Contact between intensifying screen and film | - annually
126
Speed uniformity is tested to assure that all like speed intensifying screens
Respond to the same xray stimulation by releasing the same amount of visible light - +/-10% annual
127
Darkroom fog density can be no greater than
.08 (xray) or .05 (mammo) | - semiannual
128
Sensitometry
Measuring the response of film to exposure and processing | - daily
129
Developer is sensitive to changes in
Temperature, oxidation, concentration, contamination
130
Sensitometer
Device that emits varying intensities of light in a star pattern on a film
131
Pemetrometer
Aluminum step wedge that is exposed on to radiographic film and developed
132
21 step sensitometer is
``` Most common (41% difference between steps) Also a 11 step (100% difference between steps) ```
133
Densitometer
Reads/measures image density, measures how much light is emitted to film and how much is transmitted
134
Optical density
Incident light striking the film to the intensity of light transmitters thru film
135
The higher the optical density, the less light transmitted
Therefore the greater the density
136
Density difference or contrast indicator
Found by subtracting Dmin from Dmax | +/- .15OD
137
Hyporetention (fixer = hypo)
Must be less that .05Gr/m2 - emulsion turns brown in storage due to inadequate washing of film in processor - quarterly
138
+/-2%
Collimator dial accuracy (2% of SID) SID accuracy (2% of SID) Xray to light field Positive beam limitation
139
+/-5%
Exposure reproducibility | AEC reproducibility
140
+/-10%
``` KVP accuracy mR/mAs (from installation) Exposure linearity AEC density steps Film illuminators Speed uniformity ```
141
+/-50%
Focal spot size
142
Daily
Sensitometry
143
Quarterly
Hyporetention
144
Semiannual
Collimator dial accuracy Xray to light field Positive beam limitation Darkroom fog
145
Annual
``` Filtration Half value layer Collimator illuminance SID accuracy Focal spot size KVP accuracy mR/mAs Exposure timer accuracy Exposure linearity Exposure reproducibility Protective apparel Film illuminators Film screen contact Speed uniformity ```
146
developer solution
electron soup - lots of electrons, when silver added = black metallic silver
147
the automatic processors four sections
developer fixer washer dryer
148
function of developer
convert exposed silver halide crystals into black metallic silver. latent to manifest. 95 degrees
149
function of fixer
remove unreduced silver halide crystals. | permanently "fix" image into emulsion
150
function of washer
to wash residual fixer from emulsion to improve archival quality
151
hyporetention
retention of fixer by films emulsion resulting in degrading of the stored image over time
152
dryer section
like a hair dryer, heated air blows across film to evaporate water from emulsion; emulsion is hardened and ready for handling. 130-135 degrees.
153
the basic theme to ALL digital imaging systems
to convert X-ray energy into electrons
154
analog
a device or system that represents info as continuously variable physical quantities, like a mechanical watch. X-ray film is analog.
155
digital
a device or system that represents info as continuously variable numeric values, like a digital watch. CR/DR are digital.
156
CR - Computerized Radiography
Projection Radiography which uses a PSP (photostimulable phosphor plate - IR/IP). No film.
157
A PSP is composed of
Europium Barium Fluorohalide
158
Phosphor plate function
latent image is formed by remnant radiation striking the plate. Xray is absorbed within the PSP crystal.
159
fading
the loss of latent image information over time
160
CR scanner
IR casette is read by a CR scanner which is placed in the scanner and opened up internally to remove the phosphor plate.
161
readout
CR scanner scans the IR plate with a Helium Neon light and reads the latent image fro the phosphor
162
raster pattern
side to side pattern by the laser to obtain image
163
f centers
electrons elevated out of their normal orbital rings and into higher orbitals, containing potential energy, latent image is held in this chemical code of rearranged electrons
164
"F" in F Center is for
Fabre - the German word for color
165
Latent to Manifest Image
As laser beam travels over the plate, the latent image is released as visible light, the electrons elevated to a higher orbital are "slapped" back to their normal orbit by laser. The transfer of electron energy results in the release of visible light energy.
166
Helium Neon laser
red in color
167
photomultiplier TUbe
a device that converts light energy to electrical energy
168
digitizing
the process of converting colors, shades, and shapes held as analog information into numbers
169
erasing the PSP plate for re-use
exposure to high intensity fluorescent light
170
3 inherent factors responsible for image resolution capability
image plate size laser beam size monitor matrix size
171
current resolution for CR digital imaging is
2-5 line pairs per millimeter, conventional radiography can demonstrate 6 lp/mm
172
spatial resolution
ability to see small objects | measured as lp/mm
173
Image resolution is synonymous with
image detail or image quality
174
Contrast resolution is
the number one advantage of digital imaging versus film/screen imaging
175
Quantum Detection Efficiency (QDE)
measures the efficiency of a CR system to convert remnant radiation into useful image signals
176
High QDE systems
reduce patient dose due to requiring less radiation to produce an acceptable digital signal
177
Fuji Systems
``` "s" number Inverse relationship ( ```
178
Kodak (carestream)
``` "EI" value Direct relationship (>2050=overexposed) Expected range - 1950-2050 (2000 ideal) ```
179
Agfa
``` "LGM" number Direct Relationship (>2.5=overexposed) Expected range - 1.9-2.5 (2.2 ideal) ```
180
Direct Radiography
the use of detectors that convert X-ray energy into electrical energy that is then delivered to a computer where the anatomical image is digitally processed and displayed
181
DR is a direct readout image acquisition technology.
the detector is in the flat panel and directly sends the image to a computer upon exposure. No need for processor, ADC, digitizer.
182
Steps in indirect digital image formation with Amorphous Silicon
Remnant xray beam enters Amorphous Silicon X-ray energy converted to visible light Photoconductive material receives light (converts light to electron energy) Electron energy captured by TFT device (information converted to digital signal)
183
Steps in Direct DR Formation with Amorphous Selenium
Remnant xray beam enters Amorphous Selenium and ionizes - electrons released Electrons collected by storage capacitor and transferred to TFT (information converted to digital signal)
184
Direct DR
Amorphous Selenium | xray to electron to digital
185
Indirect DR
Amorphous Silicon | xray to light to electron to digital
186
No visible light in the image formation process with ?
Direct DR
187
Thin Film Transistors
Used with both Direct and Indirect DR Collect electric charges, positioned in a matrix, detect charges on a pixel by pixel basis. Capable of very high spatial resolution (greater than 20p/mm)
188
DELS
Detector elements of a thin film transistor
189
Receiving the DR signal
Each Pixel (DEL) from the detector array (TFT) collects electrons in a capacitor, the number collected is directly proportional to the strength of the X-ray received.
190
Algorithms
A finite sequence of instructions often used for calculation and data processing. The computers mathematical recipe for processing the digital signal from the pixels.
191
Window Width
image contrast
192
Window level
image brightness/density
193
Histograms
Anatomy specific. | a graphical representation of the distribution of numerical data.
194
with a histogram the raw data is collected, then...
then interpreted and presented to the human eye
195
Look up table
a table of predetermined luminance values, a mapping function in which every pixel is changed to a new shade of gray simultaneously resulting in an image with appropriate brightness and contrast.
196
The look up table controls
automatic rescaling
197
automatic rescaling
computers ability to display an image with a diagnostically correct gray scale (within limits) Overexposed pixels are "turned down" Underexposed pixels are "turned up"
198
digital image quality factors
``` pixel/matrix size spatial resolution contrast resolution automatic rescaling histogram windowing phosphor material exposure indicator technical factors ```
199
contrast manipulation
A post processing tool used to manipulate the contrast of an image to achieve a desired image.
200
dynamic range
The range if grays that can be assigned to a pixel. | The higher the range, the more grays available to display the image.
201
image stitching
process of combining multiple images with overlapping fields of view to produce a high-resolution image
202
sampling
Process used to digitize the spatial information in an image, typically achieved by dividing an image into a square or rectangular array of sampling points.
203
shutter
The process of removing border from image as a post processing tool. Not to be used to compensate for poor collimation.
204
modulation transfer function
the ability of a system to record spatial frequencies (lp/mm) and maintain adequate contrast between adjacent line pairs
205
Nyquist Theorem
when sampling a signal, the sampling frequency must be greater than twice the bandwidth of the input signal so that image reconstruction will be nearly perfect.
206
Aliasing
When an image contains frequencies above its lp, they are under-sampled upon digitisation and appear as lower frequency information in the digital image
207
aliasing in DR occurs when
spatial frequency is greater than the Nyquist frequency | sampling occurs less than twice per cycle
208
aliasing in CR occurs when
grid errors, grid frequency must be at least 178 lpi
209
edge enhancement
computer looks for areas of common pixel signal strength across the matrix, which are then averaged and amplified to the area of interest and other tissues suppressed resulting in decreased contrast, potential for loss of detail/quantum mottle
210
smoothing
computer looks for areas of common pixel signal strength across the matrix, which are then averaged resulting in increased contrast and reduction of unwanted noise.
211
edge enhancement aka
high pass filtering | useful for viewing large anatomical structures such as organs
212
smoothing aka
low pass filtering | useful for viewing smaller anatomical structures such as trabecula
213
Signal to noise ratio (SNR)
wanted (useful) anatomical information (signal) versus the unwanted electrical interference (noise)
214
noise
detected information that adds nothing to the anatomy - quantum mottle is an example
215
spatial resolution
the ability to see small objects (lp/mm) | controlled by crystal size in CR, detector element size in DR and pixel size on a digital monitor
216
contrast resolution
Ability to differentiate between shades of gray, controlled by Bit depth, Bit range, Pixel depth
217
spatial resolution is measured in
lp/mm
218
maximum spatial resolution is the result of
a large matrix and small pixels
219
small pixels =
Better spatial resolution
220
Image matrix
contributes to spatial resolution, the layout of cells in rows and columns; the larger the matrix, the better the spatial resolution pixels, voxels, hounsfield units (CT)
221
pixel density
the number of pixels per given area
222
The human eye is capable of seeing ? shades of grey
30-32
223
the higher the contrast resolution, the more distinct adjacent structures are
with similar densities
224
digital imaging systems have significantly higher
contrast resolution than film screen
225
contrast resolution aka
pixel depth bit range dynamic range
226
contrast resolution is rated in
bit depth
227
latitude
margin for technique error
228
thousands of shades of gray
low contrast
229
Bit range
``` 1 bit = 2 shades 2 bit = 4 shades 3 bit = 8 shades 4 bit = 16 shades 5 bit = 32 shades 6 bit = 64 shades 7 bit = 128 shades etc... ```
230
Window width
selecting which of the available grays are to represented to the eye upon the monitor controls digital image contrast
231
window level
the exact center of the window selected | controls digital image brightness (density)