rad bio ch 20 Flashcards
QA deals with __
people
Program of QA monitors:
patient scheduling
reception
preparation
Quality assurance also involves image interpretation aka
outcome analysis
10 step QA program
assign responsibility
delineate scope of practice
identify aspects of care
identify outcomes that affect the aspects of care
establish limits of the scope of assessment
collect and organize data
evaluate care when outcomes are reached
take action to improve care
assess and document actions
communicate information to organization wide quality assurance programs
QC designed to ensure that the radiologist is provided with the optimal image produced through good equipment performance and resuting in minimal patient dose
dose
QC deals with ___
instrumentation and equipment
acceptable QC program consist of three?
acceptance testing
routine performance monitoring
maintaince
QC responsitly of the_____
physcist
most important patient protection characteristic of a radiographic system is ___
filtration of the beam
minimum total filtration
2.5 mm of all or equivalent
measures total filtration
HVL
filtration is measured ____
- amount
annually
-2.5 mm of al
Collimation is checked____
-tollerance
- semiannually every 6 monhs
- +_ 2% SID
focal spot frequency
-tolerance
- annually
- +_50%
calibration of KVP
-tolerancee
annually
-+_10%
exposure timer accurancy
-tolerance
- annually
- +_5% > 10ms