Which of the following will increase the amount of color displayed in the abdominal AO?
increase frame rate
decrease color gain
decrease color threshold
increase wall filter
increase frame rate
**changing to a smaller sample size will increase frame rates
What type of transducer produced the image?
sector phased array
linear sequential array
curvilinear phased array
vector phased array
sector phased array
You are evaluating an obese patient with a known AAA. While using color Doppler to assess flow within the aneurysm, minimal flow is detected in the anechoic lumen. You decrease the color scale and increase the color gain but there is no improvement. Which of the following adjustments will most improve the color display on this exam?
decrease the color threshold
switch from a 4MHz curved array to a 4MHz linear array
increase color sample size
increase the color threshold
increase the color threshold
**increasing the color threshold will increase the amount of color displayed
What determines the transducer frequency used for the evaluation of a foreign body?
composition of the foreign body
depth of the foreign body
the radiologist reading the exam
age of the foreign body
depth of the foreign body
What acoustic artifact is demonstrated by the red arrow?
reverberation
refractive shadowing
mirror imaging
ringdown
refractive shadowing
The primary factor that determines transducer selection for a biopsy is:
physician preference
thickness of the sterile cover
depth of the mass
composition of the mass
depth of the mass
Which of the following would cause artifactual debris within a simple cyst?
increased rejection settings
short distance from the transducer
activation of spatial compounding
reduced gain settings
short distance from the transducer
Which of the following best describes the areas indicated by the yellow arrows?
reverberation artifact
refractive shadowing
PE element damage
posterior shadowing
PE element damage
A large angiomyolipoma in the right kidney is most commonly associated with what acoustic artifact?
mirror image
reverberation
comet tail artifact
propagation speed artifact
propagation speed artifact
What is the BEST way to evaluate the indicated vessel for the presence of blood flow?
color Doppler
PW Doppler
power Doppler
B-color
power Doppler
**preferred for smaller vessels and those coursing perpendicular to the US beam
Which of the following will reduce ansiotropy artifact?
dual screen imaging
turn of harmonic imaging
reduce color Doppler gain
rocking the transducer
rocking the transducer
**ansiotropy artifact refers to falsely hypoechoic areas of a tendon that occur when the transducer is not completely parallel to the tendon
Which of the following acoustic artifacts is identified on the image?
reverberation
lobe artifact
mirror image
misregistration
reverberation
Overgaining on an US exam can cause ___.
blood to appear anechoic
underestimation of arterial stenosis
overestimation of Doppler velocity
increased posterior shadowing
overestimation of Doppler velocity
You are scanning the GB and there appears to be sludge within the lumen. Which of the following will aid in differentiating the sludge from artifact?
ask the patient to perform the Valsalva maneuver
increase the transducer frequency
apply color Doppler
ask the patient to fast for 3-4 more hours and rescan
increase the transducer frequency
____ occurs posterior to highly attenuating structures.
enhancement
speckle
shadowing
refraction
shadowing
Acoustic Radiation Forced Impulse (AFRI) evaluation of the liver:
is performed with a subcostal midline window
is helpful in cases of suspected abscess formation
should be performed with the patient in the prone position
should be performed in the right lobe
should be performed in the right lobe
**used to assess the liver for signs of chronic disease
All of the following are characteristics of Power Doppler, EXCEPT:
demonstrates presence or absence of flow
demonstrates organ perfusion better than color Doppler
more sensitive to slow flow than color Doppler
demonstrates velocity of flow within the organ
demonstrates velocity of flow within the organ
When evaluating a liver transplant, you must use angle correction during Doppler evaluation of:
PVs only
HAs only
HAs and PVs
HAs and HVs
HAs and PVs
What system control should be turned off when trying to differentiate a small stone in the GB from a sludge ball?
compound imaging
harmonic imaging
focal zone
time gain compensation
compound imaging
** turning this feature off can improve visualization of posterior shadowing and enhancement
If there is no diastolic flow demonstrated on a Doppler tracing of the interlobar artery of the upper pole of the right kidney, what will the RI be?
unable to calculate RI, not enough information
0
1.0
0.5
1.0
(PSV - EDV)/PSV
Which type of transducer would be used to evaluate a patient with increased LFTs and height 5’3” and weight 275lbs?
10MHz vector array
5MHz linear phased array
5MHz sector phased array
3MHz curvilinear phased array
3MHz curvilinear phased array
Harmonic imaging leads to:
increased image noise
reduced signal to noise ratio
reduced appearance of lobe artifact
improved axial but degraded lateral resolution
reduced appearance of lobe artifact
On a scrotal ultrasound exam, how do you differentiate acute epididymitis from post-vasectomy enlargement of the epididymis?
use dual screen imaging
use color Doppler
use power Doppler
use a stand off pad
use color Doppler
** hypervascularity is demonstrated with infection, but not with post-surgical changes
The artifact that is produced when the incident beam hits a curved surface and produces an acoustic shadow is referred to as:
lobe artifact
posterior shadowing
refractive shadowing
comet tail artifact
refractive shadowing