Systems Exam 3 - Excore Instrumentation Flashcards
(19 cards)
SR SHTDN FLUX HI alarm
→annunciates if flux increases to 300 cps or 5x background
→can be blocked
→this alarm also activates containment evacuation alarm
SR FLUX DBLG Alarm
→annunciates if SR flux doubles in 9 minutes
Source Range Detector
→channels N31 & N32
→BF3 Proportional Detector
→Rx Trip at 10^5 cps (1/2 detectors, can block above P-6)
→Pulse Height Discriminator: Eliminates pulses due to gamma ionization
Note: high voltage switching operations (or lightning) can spike SR detectors, causing trips.
Intermediate Range Detector
→channels N35 & N36
→Compensated Ion Chamber (gamma interactions in both chambers are cancelled out by adjusting voltage of gamma only detector (inner can)
→Rx Trip when detector current output = 25% Rx Power (1/2 detectors)
→Overcompensated (high compensation voltage) causes detector output to decrease (read lower than actual)
→Undercompensated (low compensation voltage) causes detector output to increase (read higher than actual)
Note: if one channel is undercompensated, you may not get the unblock of SR trips; can manually unblock SR trips at CB in this case
Power Range Detector
→channels N41, N42, N43, & N44
→Uncompensated Ion Chamber
→amount of gammas compared to amount of neutrons is insignificant
→amount of background gammas is directly proportional to power level
Nuclear Engineering has estimated that the effect on PR NIS is approximately 0.8 % per degree at 100% power. The effect is decreased proportionally with power. For example at 50% power, the effect is halved or 0.4% per degree.
Tech Spec 3.3.1 & 3.9.3 Source Range:
→Source Range Instrumentation is required in Mode 2 (below P6) through Mode 6
→in Mode 6, only visual count rate indication is required in the Control Room
→in Mode 3-5 the T.S. only applies if the Rod Control System is capable of withdrawal and one or more rods not fully inserted.
Tech Spec 3.3.1 Intermediate Range:
The Intermediate Range Instrumentation is required in Mode 1 (below P-10) and Mode 2 (above P-6)
1 IR Channel inoperable, Mode 1 (<P-10) & 2 (>P-6):
→reduce THERMAL POWER to <P-6 in 24 hours, or
→increase power to >P-10 in 24 hours
→Mode 2 (<P-6) - Restore to operable prior to increasing power above the P-6 setpoint
Both IR channels inoperable Mode 1 (<P-10) & 2 (>P-6):
→suspend operations involving positive reactivity additions immediately, and
→reduce power to <P-6 in 2 hours
Tech Spec 3.3.1 Power Range:
2 PR Channels Inoperable, enter LCO 3.0.3
RX > 50% PWR UP PR DET FLUX DEV HI alarm:
any one detector deviating by > 5% of the average of all upper detectors
RX > 50% PWR LWR PR DET FLUX DEV HI alarm:
any one detector deviating by > 5% of the average of all lower detectors
Power Range Deviation alarm:
deviating by >5% of the average of all detectors
From 100% power, approximately how long after a trip do we take to come down to where Source Range instruments are energized?
→approx. 15-18 minutes
→10^-10 cps
→Startup Rate = -1/3 dpm
→approx. 6 decades
Where are the overlaps between SR, IR, and PR instruments?
→0% power on PR ≈ 10^-5 amps on IR
→4 x 10^4 cps on SR ≈ 10^-10 amps on IR
Tech Spec 3.2.4 Quadrant Power Tilt Ratio (QPTR)
→action required once ≥1.02
→applicable in Mode 1, >50%
→For each 1% QPTR above 1.00, reduce power 3% (pay attention to power level)
→verify QPTR every 12 hrs
→perform evaluation to allow continued operation
→if completion time not met, reduce power to <50%
Note: this one is tricky. Action isn’t required until 1.02, but you must take the action based on the amount above 1.00
Tech Spec 3.2.3 Axial Flux Difference (AFD)
→applicable in Mode 1 >50%
→Spec not met when two or more PR NIs indicate out of spec
→30 minutes to be <50% - no time for correction
Background Info:
→AFD = ΔI
→inside the doghouse = acceptable; outside the doghouse = unacceptable
“Half power, half hour”
Note: if AFD is high on one channel, then lower detector failed; if it’s low on one channel, upper detector failed
Source Range and Level Trip Bypass Switch
Note: losing power has the same effect as pulling fuses.
Intermediate Range and Level Trip Bypass Switch
Note: losing power has the same effect as pulling fuses.
Gammametrics
→2 ranges: Low range (0.1 cps to 10^5 cps) and Wide range (10^-8 to 200% power)
→fission chamber detector
→designed to operate under accident conditions; PAM equipment
→indicates on MCB and RSP
The Power Range Summing and Level Amp is used for:
→Gain Pot for Calorimetric Adjustment
→adds upper and lower detectors together before sending values out to be used by different logics